DESCRIPTION
Status contact
OUT
Cooler
EC
P
>°
C
ON
<°
C
GL18
GL25
515
385
Powe
r
suppl
y
1
2
3
LA1.2
TRM1.2
Option
Figure 2
Design of the conditioning and sampling system SS-5 and SS-5/3
The installation of a gas cooler
and an appropriate diaphragm gas pump
will be carried out
according to the maximum flow of gas extracted (see instruction manual for individual components).
The possible combinations are listed in the following table:
SS-5...
Type of
cooler
max. gas flow [Nl/h]
Sample gas pump
SS-5
ECP1000
150
N 3 KPE
SS-5/3
ECP3000
350
N 9 KPE
The minimum amount of flow is determined by the sample gas pump (see 8.). Premature damage can
be caused to the pump membrane if less than the minimal total amount of flow is extracted as a result
of excess pressure.
All gas coolers are equipped with a Duran glass heat exchanger
. Heat exchangers of PVDF or
stainless steel are also available.
1 = Sample gas IN
2 = Sample gas OUT
3 = Condensate OUT
(all connections DN4/6)