9-3.20-MD
Gas sampling and gas conditioning technology
9
Any change in the oxygen concentration causes a linearly proportional change in the
compensation current and hence can be read directly as the oxygen value in % O
2
on a display.
Due to the very low stagnant volume (2 cm
3
) and the direct flow of the
M&C
measuring cell, an
extremely fast response time (T
90
time) of the measuring cell of 1 second can be achieved at a
high gas flow rate.
8
ASSEMBLY
In the bottom panel of the transmitter four M4 holes are provided for mounting.
A T T E N T I O N !
When mounting, ensure that no mechanical vibrations can be
transmitted to the transmitter.
Installation must be horizontal.
Illustration 2 Dimensions
View from below.
Mounting holes
Item
2
Item
3