Gas sampling and gas conditioning technology 

9-3.20-MD

 

 

12.1 

ZERO POINT CALIBRATION 

Zero point calibration of the transmitter takes place with an O

2

-free gas, such as nitrogen (N2) 5.0.  

 

  Set the zero gas flow rate with a needle valve or flow metre to a maximum of 60 l/h. The 

flow  rate  of  the  calibration  gas  should  always  be  adapted  to  the  measurement-gas  flow 
rate;  
 

  Wait for approx. 30 seconds until the display has stabilised. 

 

  If necessary, adjust the zero point with the zero point potentiometer to 0% = 4mA. The zero 

point  potentiometer  must  then  be  located  approximately  in  the  central  position.  If  the 
position  of  the  zero  point  potentiometer  deviates  grossly  from  the  centre  position,  the 
mechanical  zero  point  must  be  corrected  (see  12.1.1).  Should  this  no  longer  be  possible, 
the measuring cell is expected to be replaced. 

 
 

 

 

Illustration 4  Calibration

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

 

Содержание PMA 05

Страница 1: ...Gas sampling and gas conditioning technology 9 3 20 MD Manual Oxygen transmitter Model PMA 05...

Страница 2: ...ct addresses can be found in the appendix to this manual Please also consult our homepage www mc techgroup com for further information about our products There you will find the manuals and product da...

Страница 3: ...ning 11 12 Calibrate 11 12 1 Zero point calibration 12 12 1 1 Mechanical zero point adjustment 13 12 1 2 Interference effects 14 12 1 3 Consideration of cross sensitivities 15 12 2 End value calibrati...

Страница 4: ...when using the device Read the manual before operation and use of the equipment The instructions and warnings in the manual must be followed Work on electrical equipment may only be performed by quali...

Страница 5: ...precautions are not taken C A U T I O N without a warning triangle symbol indicates that property damage may result if proper measures are not taken A T T E N T I O N indicates that an undesirable re...

Страница 6: ...cations They are a suitable and reliable analyser unit for oxygen monitoring in different processes such as flue gas monitoring inerting systems fermentation processes process and laboratory measureme...

Страница 7: ...output 1 of span at range 3 100 digital indicator 0 1 vol O2 analogue signal output 2 of span at range 1 Sample gas outlet pressure the sample gas must flow without pressure out to atmosphere Influen...

Страница 8: ...dumbbell there is a small mirror The dumbbell is surrounded by a wire loop which is required for the compensation method The above system is fixed by a platinum strap rotationally symmetrical in a gl...

Страница 9: ...2 cm3 and the direct flow of the M C measuring cell an extremely fast response time T90 time of the measuring cell of 1 second can be achieved at a high gas flow rate 8 ASSEMBLY In the bottom panel o...

Страница 10: ...s is where the transmitter receives power and the measurement signal is transferred Terminal 11 12 Supply voltage 24V DC max 1 5 A for heating and transmitter Terminal 9 10 Temperature status contact...

Страница 11: ...ped with heating the measurement gas should be applied only after heating of the transmitter Temperature status contact closed 12 CALIBRATE Before performing calibrations the system and process specif...

Страница 12: ...e adapted to the measurement gas flow rate Wait for approx 30 seconds until the display has stabilised If necessary adjust the zero point with the zero point potentiometer to 0 4mA The zero point pote...

Страница 13: ...lows Assign zero gas as described in 13 1 Loosen fixing screw Item 4 of the photocell bracket Turn the adjustment screw Item 6 of the photocell clockwise or counter clockwise until 0 0 oxygen is indic...

Страница 14: ...2OH 2 0 78 0 88 Ethylene C2H4O2 0 54 0 60 Furan C4H4O 0 90 0 99 Helium He 0 29 0 32 n hexane C6H14 1 78 1 97 Hydrogen chloride HCL 0 31 0 34 Hydrogen fluoride HF 0 12 0 14 Hydrogen sulphide H2S 0 41 0...

Страница 15: ...tmosphere consisting of CO2 and O2 the interference effect can be eliminated easily by using carbon dioxide CO2 for zero point calibration instead of nitrogen N2 Example 2 Determining the oxygen conte...

Страница 16: ...neglect of the cross sensitivity in this example would mean a relative error of about 2 NOTICE After zero point calibration the measuring range end value must always be calibrated 12 2 END VALUE CALI...

Страница 17: ...f external contamination clean the transmitter only with a cloth dampened with soapy water 14 DECOMMISSIONING For long term shutdown it is recommended to flush the transmitter with a dry inert gas suc...

Отзывы: