The M&C liquid coolers LC-1 and LC-1S are designed with a cooling coil fully welded in the jacket tube. The LC-1S
version is equipped with a longer cooling coil.
Figure 2 Liquid cooler LC-1 with cooling coil for the liquid sample
A dial thermometer is positioned in the coolant inlet of the LC-1 and LC-1S liquid coolers (coolant IN).
In the LTC-1 liquid cooler, the liquid sample flows through a tube bundle. There is a distribution chamber in each
of the inlet and outlet areas of the tube bundle. The distribution chamber in the upper part of the tube bundle is
screwed to the liquid cooler and can be detached.
The dial thermometer of the LTC-1 liquid cooler is located in the sample outlet (see Figure 6). The dial thermometer
can be used to indicate the sample outlet temperature.
The gas and liquid coolers of the LC, LG and LT series operate with low maintenance.
The cooling effect and the stability of the cooling depend, among other things, on the coolant inlet temperature,
the coolant quantity, the sample inlet conditions, the aggregate state of the liquid sample to be cooled, ∆T
between coolant inlet temperature and sample outlet temperature as well as the ambient temperature.
The existing coolant quality and the liquid sample to be cooled must correspond with the material specification
of the cooler.
Coolant IN
Dial thermometer