M&C LC Series Скачать руководство пользователя страница 10

 

 

 

 

 

10 

 

LC, LG, LT |  1.00.01 

www.mc-techgroup.com 

8

 

Technical Data 

8.1

 

Gas Cooler 

Gas Cooler

 

LGC-1 

LGC-1S 

LGT-2 

Part No. 

04K1000 

04K1500 

04K4000 

Max. cooling capacity approx. 

900 kJ/h 

1600 kJ/h 

3600/6100 kJ/h

1)

 

Dimension cooling coil 

1 x 4/6 mm 

1 x 4/6 mm 

Cooling coil length 

3.5 m [≈ 11.5 ft] 

6 m [≈ 19.7 ft] 

Dimension tube bundle 

9 x 6/8 mm 

Tube bundle length 

0.6 m [≈ 2.0 ft] 

Connection A (Sample gas IN) 

Tube ø 6 mm 

Tube ø 6 mm 

G 3/8" i 

Connection B (Sample gas OUT)  G 1/4" i 

G 1/4“ i 

G 3/8" i 

Connection C (Condensate OUT)  G 3/8" i 

G 3/8“ i 

G 3/8" i 

Connection D/E (Coolant 

IN/OUT). 

G 1/4" i 

G 1/4“ i 

G 3/8" i 

Sample flow rate, recommended 

max. 

500 Nl/h 

500 Nl/h 

700 Nl/h 

Max. sample pressure 

10 bar 

10 bar 

10 bar 

Max. coolant pressure 

10 bar 

10 bar 

10 bar 

Coolant liquid flow rate 

50 to 300 l/h, depending on necessary cooling capacity, coolant 

temperature IN/OUT, etc. 

Differential pressure ∆P sample 

side 

30 mbar at 500 l/h 

30 mbar at 500 l/h 

< 1 mbar at 500 l/h 

Stagnant space sample side 

175 ml 

210 ml 

370/780 ml

1)

 

Max. sample inlet temperature 

300 °C [572 °F] 

300 °C [572 °F] 

300 °C [572 °F] 

Ambient temperature 

+2 to +80 °C [35.6 to 176 °F] 

Storage temperature 

-40 to +80 °C [-40 to 176 °F] 

Mounting 

Wall-mount 

Material of sample gas-

contacted parts 

Stainless steel 316Ti* 

Dimensions (W x H x D) 

110 x 400 x 125 mm 

[≈ 4.3“ x 15.8“ x 4.9“] 

110 x 700 x 125 mm 

[≈ 4.3“ x 27.6“ x 4.9“] 

120 x 750 x 125 mm 

[≈ 4.7“ x 29.5“ x 4.9“] 

Weight 

1.8 kg [≈ 4.0 lbs] 

3 kg [≈ 6.6 lbs] 

3.3 kg [≈ 7.3 lbs] 

 

 

Options for gas cooler LGT-2 

(jacket cooler with tube bundle) 

Part No. 

Pressure rating PN40 with certificate of conformity 

04K9000 

Tube bundle: tube diameter 10/12 mm

1)

 instead of 6/8 mm 

04K9010 

Connection A/B/C/D (Sample gas IN/OUT, Condensate OUT, Coolant IN/OUT):  

G 1/4“ i instead of G3/8“ i 

04K9035 

Connection A/B/C/D (Sample gas IN/OUT, Condensate OUT, Coolant IN/OUT): 

G 1/2“ i instead of G3/8“ i 

04K9040 

 

1)  = With option tube bundle diameter 10/12 mm instead of 6/8 mm. 

*  = Standard, others on request. 

 

Maximum cooling capacity refer to over-heated vapor or liquid and sufficient coolant.

 

Содержание LC Series

Страница 1: ...Series LC LG LT Gas and Liquid Cooler LGC 1 S LGT 2 LC 1 S LTC 1 Instruction Manual Version 1 00 01 Gas Cooler LGC 1 S Liquid Cooler LC 1 S...

Страница 2: ...ses in the appendix of this manual For additional information about our products and our company please go to M C s website www mc techgroup com There you will find the data sheets and manuals of our...

Страница 3: ...orage 14 10 Preparation for Installation 14 11 Mounting the Cooler 14 12 Connecting Coolant and sample to be Cooled 14 13 Commissioning 15 13 1 Setting the Temperature 15 14 Closing Down 16 15 Mainten...

Страница 4: ...e transportation correct storage as well as professional installation and maintenance work All necessary information a skilled staff will need for appropriate use of this product are given in this ope...

Страница 5: ...estionnaire The cooler must not be operated in potentially explosive atmospheres The cooler must only be operated with sample gas according to the analysis questionnaire If the cooler is used to cool...

Страница 6: ...n occur if the corresponding information is not taken into account Note Note indicates important information relating to the product or highlights parts of the documentation for special attention Qual...

Страница 7: ...ses Protect your eyes while working with chemicals or sharp objects Wear safety glasses to avoid getting something in your eyes Wear protective clothes Working with chemicals sharp objects or extremel...

Страница 8: ...Whenever you call M C regarding questions or orders for the spare parts please give us the serial number of the device 7 Description The M C gas coolers LGC 1 and LGC 1S are designed with a cooling c...

Страница 9: ...ube bundle is screwed to the liquid cooler and can be detached The dial thermometer of the LTC 1 liquid cooler is located in the sample outlet see Figure 6 The dial thermometer can be used to indicate...

Страница 10: ...l h 30 mbar at 500 l h 1 mbar at 500 l h Stagnant space sample side 175 ml 210 ml 370 780 ml1 Max sample inlet temperature 300 C 572 F 300 C 572 F 300 C 572 F Ambient temperature 2 to 80 C 35 6 to 176...

Страница 11: ...Stagnant space sample side 44 ml 76 ml 350 740 ml1 Max sample inlet temperature 300 C 572 F 300 C 572 F 300 C 572 F Ambient temperature 2 C bis 80 C 35 6 to 176 F Storage temperature 40 C bis 80 C 40...

Страница 12: ...Cooler Dimensions Figure 3 Dimensions gas cooler LGC 1 S Figure 4 Dimensions gas cooler LGT 2 A Sample gas IN B Sample gas OUT C Condensate OUT D Coolant IN E Coolant OUT A Sample gas IN B Sample gas...

Страница 13: ...8 4 Liquid Cooler Dimensions Figure 5 Dimensions liquid cooler LC 1 S Figure 6 Dimensions liquid cooler LTC 1 A Liquid sample IN B Liquid sample OUT D Coolant IN E Coolant OUT A Liquid sample IN B Liq...

Страница 14: ...ed for stationary use and if properly selected and installed guarantees long service life and minimum maintenance Shut off control devices must be provided by the customer in the inlet of the coolant...

Страница 15: ...gas carrying parts The cooling circuit should always be connected and put into operation first and then the sample circuit Caution hot surfaces possible If hot media are fed through the cooler the op...

Страница 16: ...rt gas Aggressive condensate possible Chemical burns caused by aggressive media possible For general electrical and mechanical work on the cooler wear personal protective equipment PPE in accordance w...

Страница 17: ...he maintenance cycle Warning If toxic or oxygen displacing gases are cooled with the cooler the gas paths must be purged with inert gas or air before opening the gas carrying parts Furthermore the wor...

Страница 18: ...nt to take care of the appropriate disposal of obsolete electrical and non electrical devices To help protect our environment please follow the rules and regulations of your country regarding recyclin...

Отзывы: