26
PMA 50 EX | 1.01.04
www.mc-techgroup.com
Check the analog output signals;
The signal to be measured can be calculated as follows:
(max. signal output - min. signal output)
x
Concentration [%]
Measured value =
+ min. Signal output
Full scale value [%]
A test gas concentration of 20.93 % (air) would result in the following:
Output signal
Signal to be measured with full scale value
100
%
0-10 V DC
2.09 V
Output signal
Signal to be measured with full scale value
30 %
100 %
0-20 mA
13.95 mA
4.19 mA
4-20 mA
15.16 mA
7.35 mA
Note
The mA-signal depends on the set measuring range. Therefore, it is important to
check the correctness of the selected measuring range!
Shut-off the output valve of the pressure regulator and the bottle valve and disconnect the tube
connec-
tion to the analyzer.
The full scale value calibration has been completed.
Attention
After calibration set the measuring range selection switch to desired measuring
range again.
The mA-output signal is dependent on the selected measuring range!
Содержание 1803436
Страница 41: ...www mc techgroup com PMA 50 EX 1 01 04 41 Figure 10 Circuit diagram front board PMA50 EX ...
Страница 42: ...42 PMA 50 EX 1 01 04 www mc techgroup com Figure 11 Assembly front board PMA50 EX ...
Страница 43: ...www mc techgroup com PMA 50 EX 1 01 04 43 Figure 12 Circuit diagram mains adapter PMA50 EX ...
Страница 44: ...44 PMA 50 EX 1 01 04 www mc techgroup com Figure 13 Assembly connection mains adapter PMA50 EX ...
Страница 48: ...48 PMA 50 EX 1 01 04 www mc techgroup com ...
Страница 49: ...www mc techgroup com PMA 50 EX 1 01 04 49 ...
Страница 50: ...50 PMA 50 EX 1 01 04 www mc techgroup com ...
Страница 51: ...www mc techgroup com PMA 50 EX 1 01 04 51 ...
Страница 52: ...52 PMA 50 EX 1 01 04 www mc techgroup com ...
Страница 53: ...www mc techgroup com PMA 50 EX 1 01 04 53 ...