General cells and cuvettes
Chapter 2
Zetasizer Nano accessories guide
Page 2-25
High concentration cell (ZEN1010)
The cell is supplied with the following components to prepare the sample and con-
nect the cell:
Filling the High concentration cell
Fill the cell with the prepared sample as described below. Also refer to the filling
advice given earlier in this chapter.
Filling the High concentration cell uses a similar principle to the Folded capillary
cell.
Inject the sample
slowly
until the liquid reaches the bottom of the ‘luer’ outlet
.
Check no air bubbles form in the cell. Tap the cell gently to dislodge any that
do form.
Remove the syringe and insert a cell stopper in
each
port.
Remove any liquid spilt on the electrodes.
Description
ill 8450
The High concentration cell is intended primarily for the
measurement of zeta potential of concentrated aqueous
samples. The cell can be used in conjunction with the MPT-2
for automated titrations.
The cell consists of a high precision optical measurement
block held within electrode chambers. This is all contained in
an outer cuvette sized casing assembly that allows excellent
thermal contact with the instrument cell holder.
Refer to
maintenance
and
chemical compatibility
section
later in this chapter.
Application
The High concentration cell is used for measurements of high
concentration aqueous samples.
1/32" internal bore silicon tubing with
appropriate Luer fittings
Luer plugs for manual filling
Additional fittings for connection to the
MPT-2 Titrator
Interdental brushes for cleaning of the
electrode chamber, internal flow paths
and optical block are also included
Содержание Zetasizer Nano Series
Страница 2: ......
Страница 3: ...Zetasizer Nano accessories guide MAN0487 Issue 1 1 April 2013...
Страница 12: ...Chapter 1 Introduction and accessory range Page 1 6 MAN 0487...
Страница 42: ...Chapter 2 General cells and cuvettes Page 2 30 MAN 0487...
Страница 98: ...Chapter 5 Microrheology Page 5 20 MAN 0487...
Страница 125: ......