Addendum 0090-1.0
Cleaning the cells
Cleaning the Surface zeta potential cell
Caution!
During cleaning it is vital not to let any fluid enter the top and cap area of
the cell assembly.
Any cross contamination of material from one measurement to the next could
affect the result, so it is extremely important to ensure the cell is completely clean
before use.
Cuvettes used with the Surface zeta potential cell
If a quartz cuvette was used for the measurement, it is recommended to clean
the cuvette with Hellmanex, and then rinse with copious amounts of de-ion-
ised water,
prior
to reusing it.
If a plastic disposable cell was used for the measurement, it is recommended
that this is disposed of and a new one used for all subsequent measurements.
General cleaning
As a complete assembly the cell can be cleaned using de-ionised water or with a
Hellmanex solution. If Hellmanex is used, the cell
must
be rinsed with copious
amounts of de-ionised water,
prior
to reusing it,
More efficient cleaning can be obtained by immersing the electrode and sample
holder in a gentle ultrasound bath (30 Watts) for 5 to 15 minutes.
Over time, it is likely that the electrodes will become discoloured or tarnished. This
is expected, and although it cannot be cleaned, this will not effect the quality of the
data obtained.
Cleaning the cells
- Read in addition to relevant section on page 4-5 of the
Zetasizer Nano Essentials Manual.