Maintenance and troubleshooting
Chapter 6
Visco 88
Page 6-5
Spurious or erratic results
Measuring system incorrectly installed – remove and reinstall it.
Wrong measuring system selected in the software – check the setting.
Wrong temperature displayed
If using a cup and bob, check that the temperature probe is submerged about
5mm into the sample. If the temperature shown is obviously wrong, the
temperature card may be faulty; contact Malvern Instruments.
If using a cone and plate/parallel plates, use the value displayed by the TCU
itself. (With these measuring systems the temperature probe cannot measure
the sample.)
Irregular temperature readings
These may be caused by disturbances from the mains power – check the mains
power is not affected by transient currents or varying voltages and/or frequencies.
Содержание Visco 88
Страница 1: ...Visco 88 User Manual MAN0378 Issue 1 0 October 2006...
Страница 6: ...Table of Contents Visco 88 Page iv MAN 0378...
Страница 7: ...Part 1 Operator s Guide...
Страница 8: ......
Страница 22: ...Chapter 2 What is the Visco 88 Page 2 12 MAN 0378...
Страница 34: ...Chapter 3 Making a measurement Page 3 12 MAN 0378...
Страница 40: ...Chapter 4 Other tasks Page 4 6 MAN 0378...
Страница 54: ...Chapter 6 Maintenance and troubleshooting Page 6 6 MAN 0378...
Страница 55: ...Part 2 Appendices...
Страница 56: ......
Страница 62: ...Appendix A Health and safety Page A 6 MAN 0378...
Страница 65: ...Visco 88 Page C 1 C Site requirements Appendix C of the Visco 88 User Manual MAN 0378 Issue 1 0...
Страница 69: ...Visco 88 Page D 1 D Regulatory statements EMC...
Страница 70: ...Appendix D Regulatory statements Page D 2 MAN 0378 D...
Страница 74: ...Page iv Visco 88 Index Visco 88...