32
|
5. Visual inspection of mains connection cable
Check the following points on the mains connection cable:
§
Visually inspect and check the function of the strain
relief and kink protection
§
Visual inspection of the insulation parts
§
Visual inspection of the connection cables (damage,
crushing)
§
Visual inspection of the mains connection plug
§
Visual inspection of the cable hooks
6. Visual inspection of wiring
Check the following points:
§
Damage to the cables
§
Correct cable routing
§
Proper seating of plug connections and pull-out pro-
tection
7. Visual inspection of housing
The housings must be checked for external damage and intact
seals.
8. Battery replacement
The battery must be replaced in line with specifications at least
every 2 years or after every case of emergency lowering.
9. Measurements in line with DIN EN 62353
Check the electrical components of the care bed as specified in
DIN EN 62353. Measure the leakage current with the help of the
equivalent leakage current test. The limit is ≤ 500 µA.
10. Visual inspection of grab handle for lifting pole
Check that the plastic components and straps show no signs of
damage. The grab handles must be replaced every five years.
Battery replacement
The 9 V block battery/batteries must be replaced at two-year in-
tervals and after every actuation of the electrical emergency low-
ering function in order to guarantee the functional capability of
emergency lowering. Only use brand alkali batteries and please
dispose of old batteries in an environmentally friendly manner.
The two 9 V block batteries are
inserted horizontally in the dual
drive housing.