![Malmet F105BT Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/malmet/f105bt/f105bt_operation-maintenance-and-installation-manual_3958002016.webp)
Fluid Warming Cabinet
Operation, Maintenance and Installation Manual
Issue 18
Page 15
Jun-23
4.0 Technical Specifications
4.1 Device Specifications
Capacity
–
Single
(Litres of fluids)
105 Litre models
24 Litres maximum load
210 Litre models
45 Litres maximum load
260 Litre models
60 Litres maximum load
420 Litre models
105 Litres maximum load
Shelves
–
Single
105 Litre models
(3 shelves; 2 are adjustable)
Maximum load: 8 litres per shelf
210 Litre models
(3 shelves; 2 are adjustable)
Maximum load: 15 litres per shelf
260 Litre models
(4 shelves; 3 are adjustable)
Maximum load: 15 litres per shelf
420 Litre models
(7 shelves; 6 are adjustable)
Maximum load: 15 litres per shelf
Electrical Rating
Volts
240Vac
Phase / Hz
1 ph / 50Hz
Amps
2.8 Amps
Watts
0.6 kW
Environment operating conditions
Temperature
Relative Humidity
+10
C to +25
C
+30% to +70%
Electrical Connection
IEC right angle Power Cord with
3 Pin Plug
10 Amp (into standard GPO 240
Volt)
Hot Air Fan
Continuous operation
Ball bearings 230V 50 Hz
Elements
Circular Unfinned
Incoloy 840 sheathed 10mm
diameter tubular
Rated
240V 600 watt
Control PCB
Microprocessor Control Relay
PCB (SKCM325) Access via
compatible RS232
Overtemp Protection
Electronic cutout @ 5°C above set
point
Temperature Control
Incremental 1°C
–
50°C
Factory set at 37°C
Over-temp Protection
Software
5°C above set point
Secondary Overtemp Protection
Mechanical
Auto reset Bi-Metallic Disc 60°C
Materials
Doors
Bi-parting, double glazed with
powder coated frames
Cabinet
304/4 and 430D stainless steel
Environment conditions Transport
and Storage
Fragile
Keep away from rain
Do not stack