6. 

Integrated Pneumatic Backup Test:

a.  Verify pre-use assembly flowmeter

 

set to 10 L/min

b.  Turn INOmax DS

IR

 backup switch ON

c.  Verify measured values 

NO

14-26 ppm

NO

2

≤< 

≤ 

1.0 ppm

d.  Turn backup switch OFF

7. 

Performance Test:

a. Verify O

2

 flowmeter is set to 10 L/min

b.  Set INOMAX 

dose to 

40 ppm

c. Verify 

monitored 

values 

d.  Set INOMAX dose to 0 ppm

•  “Set Dose is Zero, Please Close Cylinder Valve” 

reminder will appear- DO NOT close cylinder valve at 

this time

e.  Turn oxygen flowmeter OFF

8. 

INOblender Test:

a.  Remove injector module from pre-use assembly
b.  Remove O

2

 tubing from flowmeter and attach to 

INOblender outlet

c.  Set INOblender flow to 10 L/min, INOMAX dose to 

40 ppm

d.  Confirm acceptable NO value 

NO

32-48 ppm

e.  Set INOblender dose and flow to 0 

Pre-Use Checkout

WARNING: This reference card is not a substitute for the INOmax DS

IR

 and INOblender Operation 

and Maintenance manuals. Refer to these manuals for all applicable cautions and warnings.

Acceptable NO Value

35-45 ppm

Acceptable NO

2

 Value

<1.5 ppm

FiO

2

95% ± 3%

1.  Turn device ON, low calibration will begin and complete
2. 

Initial Connections:

  Confirm attachment of the following:

a.  Water separator cartridge, water trap bottle,  and 

patient gas sample line in place

b.  Injector module cable and tubing are connected 
c.  Plug in power cord and verify AC power light is ON
d.  Regulator to 800 ppm INOMAX cylinder
e.  Regulator hose to INOmax DS

IR

 inlet

f.  INOblender hose connected and white lock in place
g.  Oxygen source (50 psig) to back of INOblender
h.  IR cable in place

3 . Assemble pre-use set-up connectors (diagram on back 

of card). Do not turn O

2

 flowmeter ON yet.

4. 

High Pressure Leak Test:

  Open/close INOMAX 800 ppm cylinder valve

a.  Verify, at least 500 psig cylinder pressure
b.  Verify, no decrease in cylinder pressure for 

30 seconds

5. 

Manual Purge/Alarm Verification: 

a.  Press CANCEL to exit pre-use wizard.
b.  Verify the INOMAX cylinder valve is closed.
c.  Set the O2 flowmeter to 10 L/min

d.  Purge INOmax DS

IR

.

• Set the INOMAX dose to 40 ppm.
• “Cylinder Valve Closed” alarm will occur.
• Cylinder gauge pressure should drop to 0 psig.
• Measured NO2 will increase and then decrease as 

NO2 is purged from the system.

• “Low Cylinder Pressure” alarm will occur.

e.  Open the INOMAX cylinder valve.
f.  Turn the INOMAX dose to zero. The “Set Dose is 

Zero, Close Cylinder Valve” indicator will appear. 

This indicator will display anytime the set dose 

is returned to zero; however, during this pre-use 

procedure, leave the cylinder open and touch the 

screen to reset the indicator.

End of Procedure

If device is not used within 10 minutes, depressurize cylinder regulator

(The following instructions are provided for when the on-screen pre-use wizard is not used.)

2e 2f

2g

2h

2a

8d

8e

8c

8c

8a

8a

8e

8b

6b

6d

Series 3 software