6.
Integrated Pneumatic Backup Test:
a. Verify pre-use assembly flowmeter
set to 10 L/min
b. Turn INOmax DS
IR
backup switch ON
c. Verify measured values
NO
14-26 ppm
NO
2
≤<
≤
1.0 ppm
d. Turn backup switch OFF
7.
Performance Test:
a. Verify O
2
flowmeter is set to 10 L/min
b. Set INOMAX
dose to
40 ppm
c. Verify
monitored
values
d. Set INOMAX dose to 0 ppm
• “Set Dose is Zero, Please Close Cylinder Valve”
reminder will appear- DO NOT close cylinder valve at
this time
e. Turn oxygen flowmeter OFF
8.
INOblender Test:
a. Remove injector module from pre-use assembly
b. Remove O
2
tubing from flowmeter and attach to
INOblender outlet
c. Set INOblender flow to 10 L/min, INOMAX dose to
40 ppm
d. Confirm acceptable NO value
NO
32-48 ppm
e. Set INOblender dose and flow to 0
Pre-Use Checkout
WARNING: This reference card is not a substitute for the INOmax DS
IR
and INOblender Operation
and Maintenance manuals. Refer to these manuals for all applicable cautions and warnings.
Acceptable NO Value
35-45 ppm
Acceptable NO
2
Value
<1.5 ppm
FiO
2
95% ± 3%
1. Turn device ON, low calibration will begin and complete
2.
Initial Connections:
Confirm attachment of the following:
a. Water separator cartridge, water trap bottle, and
patient gas sample line in place
b. Injector module cable and tubing are connected
c. Plug in power cord and verify AC power light is ON
d. Regulator to 800 ppm INOMAX cylinder
e. Regulator hose to INOmax DS
IR
inlet
f. INOblender hose connected and white lock in place
g. Oxygen source (50 psig) to back of INOblender
h. IR cable in place
3 . Assemble pre-use set-up connectors (diagram on back
of card). Do not turn O
2
flowmeter ON yet.
4.
High Pressure Leak Test:
Open/close INOMAX 800 ppm cylinder valve
a. Verify, at least 500 psig cylinder pressure
b. Verify, no decrease in cylinder pressure for
30 seconds
5.
Manual Purge/Alarm Verification:
a. Press CANCEL to exit pre-use wizard.
b. Verify the INOMAX cylinder valve is closed.
c. Set the O2 flowmeter to 10 L/min
d. Purge INOmax DS
IR
.
• Set the INOMAX dose to 40 ppm.
• “Cylinder Valve Closed” alarm will occur.
• Cylinder gauge pressure should drop to 0 psig.
• Measured NO2 will increase and then decrease as
NO2 is purged from the system.
• “Low Cylinder Pressure” alarm will occur.
e. Open the INOMAX cylinder valve.
f. Turn the INOMAX dose to zero. The “Set Dose is
Zero, Close Cylinder Valve” indicator will appear.
This indicator will display anytime the set dose
is returned to zero; however, during this pre-use
procedure, leave the cylinder open and touch the
screen to reset the indicator.
End of Procedure
If device is not used within 10 minutes, depressurize cylinder regulator
(The following instructions are provided for when the on-screen pre-use wizard is not used.)
2e 2f
2g
2h
2a
8d
8e
8c
8c
8a
8a
8e
8b
6b
6d
Series 3 software