12

Specifications

MO15 Receiver (Rx)

Colour: red
Weight: 136g including batteries, 67g without batteries 
Dimensions: length: 91mm  width: 72mm  depth: 26mm
Battery: 3 x AA alkaline. Observe correct polarity and clean contacts when
replacing batteries.
ON/OFF: sliding switch on side
Volume: volume control wheel on the side. Turn up to increase and down to
decrease. Reset button on the top will not reset if Tx is active. Press button and
hold for 3 seconds to cancel active LED. Do not use ON/OFF button to reset the
receiver.
Green LED: flashes once every 2 seconds when Rx is ready for use. Red LED 1
to 7: ON when paired Tx is triggered. Up to 7 Txs can be tuned and paired to
the 7 LEDs. Once an LED is activated, even paired ON, when reset, it will not be
triggered again for 2 minutes. The LED and buzzer will be reactivated if the
linked Tx remains active.
Tuning point: 2 small pin holes on the back of the Rx. The one on the left is the
tuning hole. Insert the straight end of a paper clip into the hole and press
slightly then release, all the red LEDs will glow. While the red LEDs are on,
activate the Tx. The LEDs will go off and only the LED which has been selected
on the Tx will flash. When the Tx is activated again, the LED will flash and the
buzzer will sound. Green LED will flash rapidly when battery capacity is low.
Replace battery.
Default hole: the small hole on the right on the back of the Rx is the default
hole. Insert the straightened end of a paper clip and press gently then release.
This will erase all the tuned memory. Re-tuning to the Tx/s will be required.
Model: MO15 (Rx)
Standard:         and   

Sensors

Easy-Clip

©

: Length 79 cm. Keep lever up when not in use. Clean sensor regularly.

Standard:  Length 79 cm. Keep clean and wipe dry. The sensor plate and
crocodile clip MUST NOT be in direct contact with the skin. 
Bed-Mat: 54 x 42 cm. Wire length 202 cm. Keep clean and wipe dry. Always
cover with cotton sheet as well as a bath towel. Under no circumstances should
the transmitter (Tx) be attached to user’s clothing when used with a Bed-Mat.
Wander-Mat: 75 x 40 cm. Wire 150 cm.  Use pressure on or pressure off (for
pressure off use with ‘inverter’).
Dialysis-Mat: 22 x 17 cm. Wire 100 cm. Keep clean and wipe dry.

Malem Medical reserves the right to change specifications without prior notice. 
©2011 Malem Medical

6492 Master IBooklet 061010_Master IBooklet 061010  21/01/2011  14:45  Page 14

Содержание Universal MO15

Страница 1: ...m l Enuresis alarm l Sleepwalking alarm l Dialysis leakage alarm Up to seven clients can be monitored with one receiver Ideal for nursing care homes special schools hospitals institutions and individuals Helps to prevent bed sores and broken bones in Dementia Alzheimer s Memory Loss the Elderly People with Special Needs IMPORTANT Please read instruction booklet before use ...

Страница 2: ... Alarm is designed as a helpful and versatile aid It is not intended to replace human care or supervision and the manufacturer Malem Medical is not responsible for any consequences resulting from the use misuse or failure of this device ...

Страница 3: ...g Receiver Rx Sound 7 Channel Selection 8 Receiver LED Number Selection 8 Re tuning Pairing 8 Sound Selection Switch 8 Batteries 9 Guarantee 9 Repairs 9 Cleaning 10 Disposal 10 Fault Finding 10 Specification Tx Rx 11 12 Diagrams Tx Rx 13 16 How to Connect Tx with Different Sensors 17 19 The Inverter 20 21 MM100111 ...

Страница 4: ... shoulder clip is available for clinics on request if safety pin use is unsuitable iii Thread sensor lead inside top as shown in diagram above For small children loop and tie sensor lead to shorten length iv Attach sensor clip to the outside of close fitting underwear at the location where it is most likely to get wet first by inserting the folded underpants material inside the front jaw of the cl...

Страница 5: ...ants i Sandwich between two pairs of cotton briefs not boxer shorts Anchor with crocodile clip c Within Y front fold i Place sensor plate within Y fronts fold and anchor by locating the crocodile clip on the outside of the waistband of the pants Avoid nylon polyester underwear d With pull ups or nappies i Cut a small slit into the outside of the material and insert the sensor plate at a location c...

Страница 6: ...astic sheets directly under the Bed Mat as condensation will cause the alarm to trigger See diagram page 5 e Attach the Bed Mat wire to the Tx and place the Tx beside the bed See page 17 f Please tuck in the Bed Mat wire so that it is not in the way for getting in and out of bed g Upon waking to the Rx try to stop any more urine from being released and use the lavatory to completely empty the blad...

Страница 7: ...ansmitter Tx Bedside table Tuck wire from Bed Mat to alarm away neatly to prevent tripping or entanglement Tie Bed Mat around mattress Place cotton sheet over Bed Mat MALEM UNIVERSAL WIRELESS ALARM TRANSMITTER MO15 Tx Under NO circumstances should the transmitter Tx be attached to the users clothing when using a Bed Mat Holes for tying Bed Mat to the mattress Mattress Bed Mat Sheet Bath Towel ...

Страница 8: ...alarm 7 If used with Wander Mat Sensor WM a Activation by applied pressure place the Wander Mat WM in the appropriate position depending on the way you wish to use it The Wander Mat can be placed beside the bed or near the door so that treading on the Mat will activate the Tx Please place a door mat over the Wander Mat so as not to trip over the Wander Mat Plug into the Tx and place away safely Se...

Страница 9: ...Standard sensor unclip the crocodile clip and remove the sensor wipe sensor plate dry and clean regularly c If using the Bed Mat unplug the Bed Mat from the Tx remove the sheet wipe clean and dry thoroughly before repositioning d If using the pressure release Wander Mat unplug from the Tx and reposition the user on the Mat c If using the Dialysis Mat unplug from the Tx clean and dry the Mat then r...

Страница 10: ...red Rx LED number first see above b Push and release the straight end of a paper clip into the small LEFT hole at the back of the receiver All the 7 LED s will glow c Immediately and while the LED s are glowing activate the Tx you wish to tune re tune by lowering the Easy Clip sensor lever or wetting the standard sensor or Bed Mat or applying pressure on the Wander Mat or wetting the Dialysis Mat ...

Страница 11: ...Tx active when not in use as this will drain the battery Observe correct polarity and clean contacts when replacing Tx and Rx batteries Tx alkaline 23A 12V Rx alkaline 3 x AA Replace batteries on the Tx or Rx if the LEDs are rapidly flashing Safety pin Pull white cover slightly to unlock push to lock you can replace the safety pin by sliding the replacement bar to the left See page 14 Guarantee Un...

Страница 12: ...side the battery compartment are clean and free of any deposits c Sound selection switch has been wrongly set or switch No 10 has been set to OFF Reset all switches to their original position as per diagram on page 16 d Remember that the Rx will not sound again within 2 minutes of being activated by the Tx However the Rx will be activated if other Tx is activated The 2 minute delay is to allow the...

Страница 13: ... flashing rapidly a Replace battery Specifications MO15 Transmitter Tx Colour red Weight 33g including battery 25g without battery Dimensions length 50mm width 45mm depth 21mm Range approximately 25m depending on nature of the structure and the state of the battery Battery cover to remove place a slotted screwdriver in the bottom slot and lower the handle Battery alkaline 23 A Observe correct pola...

Страница 14: ...ll glow While the red LEDs are on activate the Tx The LEDs will go off and only the LED which has been selected on the Tx will flash When the Tx is activated again the LED will flash and the buzzer will sound Green LED will flash rapidly when battery capacity is low Replace battery Default hole the small hole on the right on the back of the Rx is the default hole Insert the straightened end of a p...

Страница 15: ...he battery If using the Easy Clip sensor please make sure the lever is up when not in use or unplug the sensor from the socket Malem Universal Wireless Alarm Transmitter Use Alkaline batteries ONLY MODEL MO15 Tx Sensor locking socket LED will flash rapidly for 3 minutes then turn off when the battery capacity is low Replace battery ...

Страница 16: ...o lock or unlock lock unlock Malem Universal Wireless Alarm Transmitter Use Alkaline batteries ONLY MODEL MO15 Tx ON 1 2 3 4 5 6 ON 1 2 3 4 5 6 7 8 OFF OFF Channel selection switch Receiver LED selection switch Battery compartment To open place slotted screwdriver in slot and move handle down Use 1 x 23A 12V Alkaline batteries Please observe correct polarity when replacing batteries Safety pin rep...

Страница 17: ...reless Alarm Receiver Use Alkaline batteries ONLY MODEL MO15 Rx MUTE RESET BUZZER PRESS HOLD FOR 3 SEC TO RESET Buzzer Press and hold for 3 sec to reset Mute reset Red LEDs 1 7 Each light corresponds to the transmitter it is paired with Green standby LED 2 sec flashes LED will flash rapidly when battery capacity is low Replace battery Volume Attach a label to the front of the Rx to indicate LED nu...

Страница 18: ...s and hold for 3 seconds All pairing with Tx will be lost To open battery compartment place slotted screwdriver in the slot then twist Batteries 3 x AA Please use alkaline Observe correct polarity Make sure metal contacts are clean Tx Rx Pairing Use end of paper clip Pass into hole press and release all red LEDs On activate required Tx while red LEDs are On one of the LEDs will flash the same numb...

Страница 19: ...ansmitter Tx Plug sensor into Tx socket RJ10 plug Easy Clip Sensor EC Lever Keep up when not in use Clean clip regulary Standard Sensor BT MALEM UNIVERSAL WIRELESS ALARM TRANSMITTER MO15 Tx 2 Activation using a Bed Mat RJ10 plug Malem Transmitter Tx Bed Mat BM Bed Mat connection wire Plug Bed Mat RJ10 plug into the Malem transmitter socket 54 cm 42 cm 202 cm ...

Страница 20: ...using a Dialysis Mat MALEM UNIVERSAL WIRELESS ALARM TRANSMITTER MO15 Tx RJ10 plug Malem Transmitter Tx Dialysis Mat DM DM connection wire 100 cm Plug Dialysis Mat RJ10 plug into the Malem transmitter socket 22 cm 17 cm ...

Страница 21: ...VERSAL WIRELESS ALARM TRANSMITTER MO15 Tx WANDER MAT MALEM PRESSURE MAT INVERTOR RJ10 plug Malem Transmitter Tx Wander Mat WM WM connection wire Plug inverter plug into Tx socket 5 Activation by removing pressure user must be lying in bed or sitting over WM in a chair Plug WM into inverter socket PM Inverter Please note that pressure must be applied to the Mat BEFORE attempting to set up the conne...

Страница 22: ...w If LED does not glow then change the battery Remove battery isolator if present The inverter should be used ONLY if you wish to use the Wander Mat to detect removal of pressure Please do not immerse inverter or Mat in water The RJ10 Wander Mat WM plug plugs into the inverter socket The RJ10 plugs into the transmitter Tx Battery test button Use Alkaline batteries ONLY LED light ...

Страница 23: ...t polarity Contacts clean and battery not loose inside compartment Pull metal contact to make good connection Use only 23A 12V Alkaline battery RJ10 socket RJ10 plug Battery test button Use Alkaline batteries ONLY Use slotted screwdriver to open Place tip in slot and twist ...

Страница 24: ...t 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a pa...

Отзывы: