7. MEASURING TEMPERATURES
7.1. CONNECTING A THERMOCOUPLE

To change the thermocouple type, see “Changing Setup Options."                      
The North American ANSI Color Code is:

1.  Plug a thermocouple into the input connector(s). (Make sure that the 

polarity is correct.)

2.  Set the thermometer for the correct thermocouple type.

7.2. DISPLAYING TEMPERATURES

1.  Press     to select the correct temperature scale.
2.  Hold or attach the thermocouple(s) to the measurement location. The 

temperature reading appears in the primary display.

3.  Press     to toggle between showing the T1, T2, , and T1-T2 readings in 

the primary or secondary display.

Notes: 

The display shows "- - - -" when a thermocouple is not connected.

The display shows OL (overload) when the temperature being measured is 
outside the thermocouple's valid range.

7.3. HOLDING THE DISPLAYED READINGS

1.  Press     to freeze the readings on the display. The display shows HOLD.
2.  Press     to toggle showing the T1, T2, or T1-T2 readings in the primary or 

secondary display.

3.  Press     again to turn off the HOLD function.

7.4. VIEWING THE MIN, MAX, AND AVG READINGS

1.  Press     to step through the maximum (MAX), minimum (MIN), or the 

average (AVG) readings. The elapsed time since entering MIN MAX mode, 
or the time at which the minimum or maximum occurred appears on the 
display.

2.  Press     button for 3 seconds to exit MAX, MIN and AVG mode.

7.5. USING THE OFFSET TO ADJUST FOR PROBE ERRORS

Use the offset option in Setup to adjust the thermometer's readings to 
compensate for the errors of a specific the thermocouple and IR temperature. 
The allowable adjustment range is ± 5.0°C or ± 9.0°F.
1.  Plug the thermocouple into the input connector.
2.  Place the thermocouple in a known, stable temperature environment (such 

as an ice dry well calibrator).

3.  Allow the readings to stabilize.
4.  In Setup change the offset until the primary reading matches the 

calibration temperature. ("Changing Setup Options.")

7.6. REPLACING THE BATTERIES

1.  Turn off the thermometer if necessary.
2.  Loosen the screw and remove the battery door.
3.  Replace 1.5V batteries.
4.  Replace the battery door and tighten the screw.

Type    

J  

K

Color   

Color  

Yellow

T1-T2

T1-T2

UNITS