4.2. USING THE THERMOCOUPLE (S)

1.  Plug the therrnocouple(s) into the T1 or T2 input connector(s).
2.  Press     to turn on the thermometer. After 1 second the thermometer 

displays the first reading. If no thermocouple is plugged into the 
selected input or the thermocouple is "open," the display shows"- - - -".

4.3. CHANGING SETUP OPTIONS

Use Setup to change offset, sleep mode settings. The thermometer stores 
the settings in its memory.

4.4. SETUP OPTIONS

Option  

Menu item  

Settings

Type  

TYPE  

K or J

Offset  

OFFSET  

T1 or T2

Sleep Mode   SLP   

(sleep mode on) or OFF 

(sleep mode off)

4.5. ENTERING OR EXITING SETUP

When the thermometer is in Setup mode, the display shows SETUP.    
Press     to start or exit Setup.

4.6. CHANGING A SETUP OPTION

1.  Press     or     to scroll to the setup option you want to change.
2.  Press     to indicate that you want to change this setting.
3.  Press     or     until the setting you want to use appears on the display.
4.  Press     to store the new setting in memory.

Notes:

 Setup is disabled in 

MAX MIN

 mode .

5. OFFSET

The primary display shows the temperature plus the offset and the
secondary display shows the offset. You can store individual offsets
for T1, T2.

6. AUTO POWER OFF MODE

The thermometer enters sleep mode (default). That is to say, the meter 
will automatically shut off after 20 minutes if no button press occurs for 20 
minutes. When the thermometer is in Setup mode, the display shows 
SETUP. Press     or     to scroll to the"SLP" page . Press     to indicate "ON" 
or “OFF". Press     or     until the setting you want to use appears on the 
display (fig.2). Press     to store the new setting in memory. ON (sleep 
mode on) or OFF (sleep mode off).

T1-T2

T1-T2

T1-T2

T1-T2

Отзывы: