• In most cases, the reflected temperature is identical to the ambient air
temperature.
• Only when objects with strong emissions with much higher temperature
are in the proximity of the object being measured should be determined
and used.
• The reflected temperature has only little effect on objects with high
emissivity.
• The reflected temperature can be set individually.
• Follow these steps to get the right value for the reflected temperature.
1. Set the emissivity to 1.0.
2. Adjust the optical lens to near focus.
3. Looking in the opposite direction away from the object, take a
measurement and freeze the image.
4. Determine the average value of the image and use that value for your
input of reflected temperature.
6.11. Thermal Imager Reporter Software
• Thermal Imager Reporter software is supplied with the Thermal Imager.
• This Software is intended for Thermal Imager and contains feature to
analyze images, organize data and information, and make professional
reports.
• Thermal Imager Reporter software allows audio annotations and
commentary to be reviewed on a PC.
7.Menus
The menus, together with buttons, are access for image measurement,
Emiss, Palette, temperature measurement range, take photos and videos,
review, and settings.
7.1. Main Menu
• Press "
Menu/OK
" Button or touch the screen, the main menu will pop
up.
• Main Menu is the main interface of the Thermal Imager's menus.
• It contains five items such as Measure parameters, Measure tools, Image
mode, Palette, system Settings.