7
5-9.999Hz
99.99Hz
±(1.5% r 6 digits)
999.9Hz
9.999KHz
Range & Resolution
4.8. Frequency Sensitivity: >8Vrms
Accuracy (% of r digits)
4.9. Temp (type-K) (probe accuracy not included)
-20.0 to 1000°C
±(3%rdg + 3°C)
-4.0 to 1800.0°F
±(3%rdg + 6°F)
Range & Resolution
Accuracy (% of r digits)
Clamp size
Opening 1.3" (33mm) approx
TRMS
The AV voltage and AC current of this meter is
measured by True RMS. TRMS measurement is
different from mean measurement. The mean
measurement method can only measure the
symmetric waveform, such as sine wave. TRMS
measurements can reliably measure any
irregular waveform and obtain valid values for
AC voltage or AC current.
Diode Test
Test current of 0.3mA typical; Open circuit
voltage 3.2V DC typical.
Continuity Check
Threshold <50Ω; Test current < 0.5mA
Low Battery Indication " " is displayed
Range
Basic Functions
4.11. General Specifications
-30.0 to 350.0°C
±(2%rdg + 2°C)
-22.0 to 662.0°F
±(2%rdg + 4°F)
Range & Resolution
4.10. IR Temp (D:S=4:1)
Accuracy (% of r digits)
Note:
Accuracy is given as ±(% of counts of least significant digit)
at 23
°
C±5°C, with relative humidity less than 80%RH. The precision index
mentioned above refers to the accuracy of 40%~60% range of each
measurement range. The accuracy index of the current measured outside
the measuring range increased by 1.5%, and the accuracy of other
measurement functions increased by 1%. Check waveform is sine wave.
Current accuracy assessment shall be based on the position of clamp center.
Содержание MT767
Страница 1: ...MT767 600A AC DC CLAMP METER WITH IR THERMOMETER INSTRUCTION MANUAL...
Страница 2: ...2...