with the flask swirling once every minute. The solution should be heated until 

all crystals are dissolved. This is best viewed against a light background. 

Crystals appear as translucent crystals. These will interfere with sample 

migration if not completely dissolved. 

 

Note:

 The gel must be cooled to between 50°C and 60°C degrees before 

pouring. 

 

3. Install the Gel Tray in the Gel Maker Stand. 

4. Select the suitable Analytical Comb for the application.       

5. Pour approx. 25 to 30 ml of agarose solution for a large tray; pour 12 to 15 

ml for a small tray. The tray contains the agarose level for reference. 

 

Note:

 Pour the agarose carefully so as not to generate bubbles. Any bubble 

that surfaces could be smoothed to the edge of the gel and dispersed 

using a pipette tip. 

 

 

6. Place the comb(s) into the grooves. 

7. As for the MJ-105 series, both Gel Maker Stands can be connected together 

for user convenience. 

 

Gel Running 

 

1. Carefully remove the comb and transfer the gel, including the tray, to the 

electrophoresis tank. 

2. Refer to the molecular biology protocols for sample preparation. 

3. Fill the electrophoresis tank with buffer until the gel is just filled to the top. 

 

DOMINIQUE 

DUTSCHER 

SAS

Содержание MJ-105-S

Страница 1: ...Mini Horizontal Electrophoresis System Instruction Manual Catalog No MJ 105 S MJ 105 R MT 108 Version 01E Date 2014 3 20 D O M I N I Q U E D U T S C H E R S A S...

Страница 2: ...i Electrophoresis Tank Cover 1x Gel Maker Stand 1x 105 x 83 mm Tray 2x 50 x 83 mm Tray 1x Double Side Comb 6 teeth x 2 12 teeth Prep 1 Marker 2 Prep2 Marker 2 23 teeth 1x Double Side Comb 8 teeth x 2...

Страница 3: ...scription 6 Section 2 Technical Specifications 8 Section 3 Installation Instructions 9 Section 4 Operation Instructions 10 4 1 Controls and Features 10 4 2 Start the operation 10 Section 5 Maintenance...

Страница 4: ...is strongly recommended for the user to read the following points carefully before operating this equipment 1 Read and follow the manual instructions carefully 2 Do not alter the equipment Failure to...

Страница 5: ...Environmental Conditions Ensure that the instrument is installed and operated strictly under the following conditions 1 Indoor use only 2 95 RH 3 75 kPa 106 kPa 4 Altitude must not exceed 2000 meters...

Страница 6: ...rs should check with the manufacturer s instruction to see if the proposed method will damage the equipment Equipment Operation Follow the guidelines below to ensure safe operation of the unit 1 NEVER...

Страница 7: ...or very fast and clear separation of DNA or RNA restriction fragments This series is offered as a complete system including all the necessary accessories for the user to cast run and perform electroph...

Страница 8: ...of agarose gel in one time for MT 108 High temperature capability 130 C Easy sample loading Great indications for gel making and running No tapes clamps or springs needed D O M I N I Q U E D U T S C...

Страница 9: ...PC Polycarbonate Construction Material Transparent Lid AS Acrylonitrile Styrene Copolymer Temperature Capacity 130 C Unit Dimension Tank and Transparent Lid 140 x 140 x 53 mm W x L x H Gel Dimension T...

Страница 10: ...Black well visualization strip Migration distance index line Rapid Casting Gel Use gel maker stand Operating Temperature 4 40 C Weight 0 7 kg Section 3 Installation Instructions 1 Seat the tray in th...

Страница 11: ...esired ratio 2 Dissolve the agarose powder by heating the agarose either on a magnetic hot plate with a stirring bar or in a microwave oven If using the microwave method the microwave should be set at...

Страница 12: ...for a large tray pour 12 to 15 ml for a small tray The tray contains the agarose level for reference Note Pour the agarose carefully so as not to generate bubbles Any bubble that surfaces could be smo...

Страница 13: ...ly connect the power cords to the power supply unit 5 Turn on the power supply unit and adjust to the desired electrophoresis running conditions 6 When the electrophoresis running is complete turn the...

Страница 14: ...tes After that rinse with distilled water Air dry before use Some cleaning agents are not supposed to come in contact with the electrophoresis systems such as Acetone Phenol Chloroform Alcohol and Alk...

Страница 15: ...phoresis Packages MJSS Mini300 Package of 2 x MJ 105 S and Mini 300 MJRR Mini300 Package of 2 x MJ 105 R and Mini 300 MTT Mini300 Package of 2 x MT 108 and Mini 300 MJSR Mini300 Package of MJ 105 S MJ...

Страница 16: ...01 MT CASTER C contains 1 x MT G01 1 x MT T01 2 x MT 02 1 x MT C03 and 1 x MJ C01 Single Electrophoresis Packages MJS Mini300 MJR Mini300 MT Mini300 Dual Electrophoresis Packages MJSS Mini300 MJRR Min...

Страница 17: ...the warranty is limited to the receipt of reasonable proof by the customer that the defect is embraced within the terms of the warranty All claims made under this warranty must be presented to Major S...

Страница 18: ...17 MEMO D O M I N I Q U E D U T S C H E R S A S...

Страница 19: ...18 MEMO D O M I N I Q U E D U T S C H E R S A S...

Страница 20: ...19 MEMO D O M I N I Q U E D U T S C H E R S A S...

Отзывы: