Operation Manual touchTymp MI 24 and MI 34 Version
8107420 Rev. 6
26
03/09/2018
The Shoulder Box
Use the clothing clip on the Shoulder Box to secure the probe to clothing or bedding
and insert the probe gently into the ear of the subject.
Figure 21
Figure 22
Figure 23
Contralateral headphone with the Shoulder Box
An additional jack on the Shoulder Box allows connection
of the Contralateral headphone (3.5 mm jack).
NOTE:
The 6.3 mm Contralateral headphone jack on the
back of the device can be used with the Pen Probe or the
Shoulder Box (see Figure 16, plug 9).
The Indication Light
The indication light displays the different states of the measurement by color and the
presentation modus (flashing/continuous). Table 6 gives explanation to the different
indications.
Table 6 Indication Light
PROBE COLOR
EXPLANATION
Red
Right ear is selected. Probe is out of ear.
Blue
Left ear is selected. Probe is out of ear.
Green
Probe is in the ear and is sealing, test is running or done.
Yellow
Probe is in the ear and blocked or leaking. If the indicator
remains “yellow” (sealing), the screener must improve the
position of the probe in the ear:
1. Reinsert the probe for better placement.
2. Inspect the probe tip for any blockage.
3. Verify eartip has the correct size, new eartip may be required.
White
An error has occurred. Confirm connection of probe and/or
restart the device.