Operating Instructions MA 53
Geba_MA53_e_13a.doc
25 851 430/3 11/11
5 Speech audiometry
5.1 Connecting the speech source
5.1.1 Connecting a CD-player
To conduct speech tests using speech test
material you can connect a CD-player to the
MA 53. Use a standard phono plug cable to
connect the right and left outputs of the
device with the corresponding inputs
and
on the rear panel of the MA 53. To avoid
pinching the cables when closing the cover,
lay the cables in the slots provided.
If you are using a small portable CD-player
or tape you can place it on top of the cover
of the MA 53. This saves space and makes the usage of the CD-player or tape
very convenient.
Caution: If you are using a CD- or tape-player powered by electrical current,
the player must meet electrical safety requirements, such as IEC 601-1 or UL.
This is to avoid electrical shock of either the patient or you. If you are not sure
if your player meets these requirements it is safer to use battery power.
5.1.2 Connecting a Microphone
Connect the microphone to the SPEECH
MIC connector
on the rear panel of the
MA 53 (see picture 30). To avoid pinching
the cables when closing the cover, lay the
cables in the slots provided. Place the
microphone on top of the cover of the MA
53. The microphone can be used for live
speech audiometry and communication
with the patient. If your instrument is
supplied with a monitor/mic headset,
connect the phone plug labeled M with the
SPEECH MIC connector
.
Picture 29 MA 53 rear panel
with TAPE/CD input
Picture 30 MA 53 rear panel
with SPEECH MIC input