4.5.1.2 The DLI (Lüscher) Test
The Intensity
D
ifference
LI
men for loudness test (or
J
ust
N
oticeable
D
ifference,
JND
) is another recruitment test. A pure tone is amplitude
modulated at a rate of 2/second. The patient has to determine whether this
signal was steady or undulating (beating).
To start the test select a modulation
amplitude from 0.5 dB to 4.0 dB in the
suprathreshold menu (as decribed before
under 4.5.1) and press the STIM bar
or
↔.
The LCD-display
shows DLI, along with the chosen modulation amplitude
(see picture 18).
Now adjust the required test level with the level control knob
or
↕ and
present the test signal by pressing the STIM bar
or
↔.
To exit the test press the PULSE buttom
longer than 2 seconds. The
suprathreshold tests menu is now displayed. To exit the suprathreshold test
menu briefly press the FM button
.
Picture 18 MA 53 display with DLI test