Operating Instructions MA 50
Geba_MA50_Fe_11a.doc
5 851 406/2 11/11
3 Getting started
3.1 Unpacking your instrument
Your MA 50 was carefully inspected and packed for shipping. However, it is
a good practice to thoroughly inspect the outside of the shipping box for
signs of damage. If any damage is noted, please notify the carrier
immediately.
Carefully remove the instrument from the shipping box. Remove the plastic
bag from the instrument and inspect the case for any damage.
Notify the carrier immediately if any mechanical damage is noted. This will
assure that a proper claim is made. Save all packing material so the claim
adjuster can inspect it as well. Notify your dealer or MAICO when the
adjuster has completed the inspection.
SAVE ALL THE ORIGINAL PACKING MATERIAL AND THE SHIPPING
CONTAINER SO THE INSTRUMENT CAN BE PROPERLY PACKED IF IT NEEDS
TO BE RETURNED FOR SERVICE OR CALIBRATION.
All accessories are already packaged in the compartment connected with the
MA 50. Please check that all accessories listed below are received in good
condition. If any accessories are missing or damaged, immediately notify your
dealer or MAICO.
Standard accessories:
-
1 Headphone TDH 39
-
1 Bone conduction receiver B 71 with headband
-
1 Patient response switch
-
1 Power cable
-
1 Audiogram form (50 sheets)
Calibration of the device:
The instrument, headphones, bone conduction receiver as well as the
optional insert phone match one another and have the same serial number
(e.g. 6631520). Because they have been calibrated with this particular
instrument, use of other transducers is not allowed. If one of the acoustic
transducers is replaced, the instrument must be recalibrated.
The use of non-calibrated audiometers leads to incorrect measurements!