9 Trouble shooting
If you should find that your instrument is no
longer working properly during a test run, please
check the following points:
Diodes in the buttons do not light up and
there is nothing to see on the LED-display
:
-
Is the power cord plugged in correctly?
-
Check the power fuses
♦.
Disconnect the power plug, unscrew the fuses
♦
alongside the power cord socket and check the
fuses. If the wire in the glass vial is broken, please
insert new fuses (1 A slow blow).
Diodes light up, but test tone is absent:
-
Green STIM LED
or
does not light:
-
Press STIM bar
or ↔
(When the diode lights up and the tone is
available, the device works in presenter mode).
-
Green STIM LED
or
does light:
-
Is level control
or
↕ set to an audible level position?
-
Are the acoustic receivers connected to the correct socket?
-
Is the required receiver with button
or
selected - is the correct LED
lit?
-
Is the lead connecting the headphones or the bone conductor loose
or defective?
If your instrument still does not work properly after this short check, please
consult your dealer or service center.
Picture 21 power socket and