Operation Manual MA 41
8100456 Rev.9
67
01/16
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The MAICO MA41 is intended for use in the electromagnetic environment specified below. The customer
or the user of the MA41 should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance
Electromagnetic
Environment-Guidance
Electrostatic Discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material the
relative humidity should
be greater than 30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
±2 kV for power supply
lines
±1 kV for input/output
lines
Mains power quality
should be that of a
typical commercial,
hospital, or residential
environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Mains power quality
should be that of a
typical commercial,
hospital, or residential
environment.
Voltage Dips, Short
Interruptions and
Voltage Variations on
Power Supply Lines
IEC 61000-4-11
<5%
U
T
(>95% dip in
U
T) for
0.5 cycle
40%
U
T
(60% dip in
U
T) for 5
cycles
70%
U
T
(30% dip in
U
T) for 25
cycles
5%
U
T
(>95% dip in
U
T) for 5
sec
<5%
U
T
(>95% dip in
U
T) for
0.5 cycle
40%
U
T
(60% dip in
U
T) for 5
cycles
70%
U
T
(30% dip in
U
T) for 25
cycles
5%
U
T
(>95% dip in
U
T) for 5
sec
Mains power quality
should be that of a
typical commercial,
hospital, or residential
environment. If the user
of the MAICO MA41
requires continued
operation during power
mains interruptions, it is
recommended that the
MA41 be powered from
an uninterrupted power
supply.
Power Frequency (50/60
Hz)
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Note:
U
T is the a.c. mains voltage prior to application of the test level.