7.2.3 Speech Audiometry
To select speech audiometry, select Speech (16) and the test SRT or WRS (13).
Use the frequency up (4) and down (5) key to select a word in the word list.
For the WRS test the display shows the percentage of correctly repeated words.
To count the word as correct press the frequency key (4) up.
To count the word as incorrect press the frequency key (5) down.
To clear the counter, press the Reset key (7) on the left side of the display.
Speech Calibration
To calibrate the MIC or CD speech input, select the input with the SIGNAL
selector key (11). Press the Monitor button (6), so that the monitor setting
screen appears. Now press the InCal button (17) on the right side of the display
and the calibration screen appears.
Play the reference signal with the CD player or speak into the microphone. Use
the left or right intensity dial (1) and adjust the levels until both VU-meters show
all yellow and one green light. If one or more red lights are displayed, reduce
the level using the corresponding dial (1).
Store the calibration and leave the calibration mode by pressing the OK (6)
button on the left side of the display.
7.2.4 Documentation of the Results
Printout
Enter the user Menu by pressing the function button (17) and press the Print
(13) button to printout the results.
Store Results as a PDF
Enter the user Menu by pressing the function button (17) and press the button
PDF (13) to save the results as PDF file on SD memory card or USB flash drive.
Store the Result in the Patient List
Enter the User Menu by pressing the function button (17) and press the button
Patient (15) to enter the patient list. Enter a new patient by pressing the button
New Patient (15). It is also possible to print a result from the patient list or to
create a PDF of the result. Select a stored session of a patient in the patient list
and press then the button PDF (11) or Print (13). In this case, the patient
information will be included in the PDF or printout. Press the button New (7) in
the patient list to start a new session.