Maico MA 1 Скачать руководство пользователя страница 8

 

MA 1 Operating Manual 

 

4                                                 

8100216-3

 (D-0100192) 

 

3.1.

 

Important Safety Note  

The MA 1 should always be operated in a quiet room with minimal magnetic influence, 
to ensure that examinations are not disturbed by external noise.  

Electro-medical  instruments  that  emit  strong  electromagnetic  fields  (e.g.  microwaves, 
radiotherapy devices) can affect the operation of the MA 1   

Therefore, the operation of these instruments in close proximity to the MA 1 should be 
avoided at all times.  

The  examination  room  should  have  a  normal  temperature  between  15˚C/  59˚F  and 
35˚C/ 95˚F.  If the instrument has cooled down during transportation, please wait for it 
to warm up to room temperature before operation. 

Attention 

PLEASE READ THE ENTIRE MANUAL CAREFULLY BEFORE OPERATING THIS 
INSTRUMENT. 

Please only use this instrument as described in the manual. 

Please familiarize yourself with the instrument and its operation before using.  

Should defects or damages be suspected, please do not, under any circumstances, use 
or attempt to fix the instrument yourself. 

Calibration  of  the  instrument:  The  audiometer  and  the  headphone  complement  each 
other  and  share  the  same  serial  number  (i.e.  63252).  Therefore,  the  instrument  shall 
not be used with any other headphone prior to recalibration. Recalibration also needs 
to be conducted, when a defective headphone is replaced. 

 

NOTE: Uncalibrated instruments may lead to faulty measurements.

 

Take note to ensure that all the accessories have been properly connected.    

To  avoid  person-to-person  cross  contamination  of  communicable  diseases,  parts  that 

come in direct contact with the patient (i.e. earphone cushions) should be 
disinfected using commercial disinfectant after each use.  

In accordance with the Electronic Equipment Act for disposal of electronic 
equipment, the customer is obliged to dispose of the used consumables, 
according to appropriate regulation at own cost. 

This  equipment  is  intended  to  be  connected  to  other  equipment  thus  forming  a 
Medical Electrical System. External equipment intended for connection to signal input, 
signal output or other connectors shall comply with the relevant product standard e.g. 
IEC  60950-1  for  IT  equipment  and  the  IEC  60601-series  for  medical  electrical 
equipment.  In  addition,  all  such  combinations  –  Medical  Electrical  Systems  –  shall 
comply with the safety requirements stated the general standard IEC 60601-1, edition 

Содержание MA 1

Страница 1: ...MAICO Diagnostics 10393 West 70th Street Eden Prairie MN 55344 USA Toll Free 888 941 4201 MA 1 Operating Manual...

Страница 2: ......

Страница 3: ...ions 6 5 Battery Operation 6 6 Front Panel Controls and Display 7 7 Audiometric Testing 8 7 1 Instructing the Patient 8 7 2 Test Environment Ambient Noise 8 7 3 Conducting the Test 9 8 Warranty and Di...

Страница 4: ......

Страница 5: ...info maico diagnostics com Internet www maico diagnostics com Copyright 2015 MAICO Diagnostics All rights reserved No part of this publication may be reproduced or transmitted in any form or by any me...

Страница 6: ...1 to ensure its user friendliness meaning that its operation is simple easy to learn and to understand The MA 1 is an ultra portable audiometer and a perfect choice for on the go hearing professionals...

Страница 7: ...s or schools 3 Description The MA 1 screening audiometer is designed to be a device for screening for hearing loss Output and specificity of this type of device are based on the test characteristics d...

Страница 8: ...headphone complement each other and share the same serial number i e 63252 Therefore the instrument shall not be used with any other headphone prior to recalibration Recalibration also needs to be con...

Страница 9: ...horoughly inspect the exterior of the shipping box for any sign of damage or tampering Should any damage be noted please notify the carrier immediately If the content box has been damaged during trans...

Страница 10: ...teries 1 Pull the tab on the battery compartment and remove panel 2 Insert 2 AA batteries oriented as shown 3 Reposition the panel and snap the tab back into place Battery Life Maico Recommends using...

Страница 11: ...3 D 0100192 6 Front Panel Controls and Display Figure 3 1 Power button and R L Ear Selector 2 LED Ear Indicator 3 LED Level Indicator 4 dB Level selector buttons 5 Frequency selector buttons stimulus...

Страница 12: ...placement of the earphone cushion on the ear i e hair earrings eyeglasses hearing aids etc b Adjust the headband so the earphone cushions are centered over the ears and the head The earphone cushions...

Страница 13: ...atient responds Once the patient responds lower the tone by 10dB until he she does not respond Continue in this down 10dB up 5dB pattern until a threshold is established or screening protocol has been...

Страница 14: ...o so In the event of repair during the guarantee period please enclose evidence of purchase with the instrument In order to ensure that your instrument works properly the audiometer should be checked...

Страница 15: ...ly contaminated it is strongly recommended to remove them from the transducer before they are cleaned with a disinfectant The use of organic solvents and aromatic oils must be avoided Always disconnec...

Страница 16: ...provided with the instrument 2 Conduct similar tests with several young adults on whom you will be able to make subsequent retests recording these on audiogram cards also 3 File these audiogram cards...

Страница 17: ...t an annual inspection and calibration should be performed The annual check ups can be performed by one of MAICO s authorized service centers 11 2 Instrument Handling The instrument should be checked...

Страница 18: ...m at full intensity Rise fall Times 35 msec Calibration Air Conduction ISO 389 1 ANSI S3 6 Warm Up Time None Dimensions W x D x H 6 35 x 15 24 x 2 22cm 2 5 x 6 x 7 8 inches Weight 2 1 lbs with accesso...

Страница 19: ...acture yyyy Registration Label Address of Manufacturer Registration Label Package Label IFU Return to Authorized Representative Registration Label B Patient Applied Part according to IEC60601 1 Regist...

Страница 20: ...ons are subject to change without notice MAICO Diagnostics 10393 West 70th Street Eden Prairie MN 55344 Phone 1 888 941 4201 Fax 952 903 4100 e mail info maico diagnostics com Internet www maico diagn...

Страница 21: ......

Страница 22: ...8100216 3 D 0100192 06 15...

Отзывы: