Mahr GmbH,  Millimar 1901/TA

5

Order No.

Last Modification

Version

3761004

20.08.2018

----

Translation of the original operating instructions

Contents

Permitted Uses ................................................1

Disposal .............................................................2

Voided Warranty ............................................2

Safety Instructions .........................................3

1  Introduction .......................................6

2  Operating Elements ........................6

3  Working with 1901/TA ...................7

3.1  Zero adjustment ..........................................................7

3.2  Adjusting the 1901/TA amplification .............8

3.3  Further settings on the PCB ................................9

4  Technical Data ................................ 10

5  Contacts at Mahr .......................... 11

6  Warranty .......................................... 11

Содержание Millimar 1901/TA

Страница 1: ...Mahr GmbH Standort Esslingen Reutlinger Str 48 73728 Esslingen Betriebsanleitung Operating Instructions Millimar 1901 TA...

Страница 2: ...eser Betriebsanleitung finden Sie folgende Zeichen Allgemeiner Hinweis Wichtiger Hinweis Nichtbeachtung kann zu fehlerhaften Ergebnissen oder Sch den am Ger t f h ren i Bestimmungsgem e Verwendung Der...

Страница 3: ...hen Gesundheit Elektrische Altger te die nach dem 23 M rz 2006 durch uns in den Verkehr gebracht wurden k nnen an uns zur ckgegeben werden Wir f hren diese Ger te einer umweltgerechten Entsorgung zu D...

Страница 4: ...ng und jeder Eingriff in den Verst rker darf nur mit ausdr cklicher schriftlicher Genehmigung der Mahr GmbH durch entsprechend ausgebildetes und geschultes Fachpersonal erfolgen Unerlaubtes ffnen des...

Страница 5: ...rwendung 1 Entsorgung 2 Gew hrleistungsverlust 2 Sicherheitshinweise 3 1 Einleitung 6 2 Bedienelemente 6 3 Arbeiten mit dem 1901 TA 7 3 1 Nullpunkt abgleichen 7 3 2 Verst rkung des 1901 TA abgleichen...

Страница 6: ...usgang darf mit minimal 5 kOhm belastet werden der Stromausgang darf mit max 2 kOhm belastet werden Der Spannungsausgang kann bei Leitungsl ngen bis 10 m verwendet werden wobei bei Kabell ngen ber 2 m...

Страница 7: ...erden so dass er ein Signal nahe null liefert Zus tzlich sollte der Tr gerfrequenzverst rker auf den Taster nullpunkt abgeglichen werden Der Nullpunktsteller hat einen Einstell bereich von ca 100 m un...

Страница 8: ...r das in Abb 3 mit Verst rkung bezeichnet wird F r den Abgleich sind zwei kalibrierte Endma e als Ma verk rperung erforderlich Der Abgleich der Verst rkung erfolgt unter Spannung am offenen Ger t und...

Страница 9: ...pannungsfrei geschalteten Ger t F r das Umstecken der Br cken Jumper ist eine kleine Pinzette erforderlich Messbereich ndern Messbereich BR 1 BR2 125 m 1 2 1 2 250 m 3 4 1 2 500 m 5 6 1 2 1000 m 7 8 1...

Страница 10: ...43 x100 mit Spannblech 170 x 43 x 100 Befestigungsabstand 149 mm Befestigungsschrauben M5 Kopf 10 mm Messbereiche 125 m einstellbar 250 m 500 m 1 000 m 2 000 m Ausgangsspannung 5 V einstellbar 10 V St...

Страница 11: ...den nat rlichen Verschlei sowie M ngel die durch unsachgem e Behandlung bestimmungswidrige Verwendung oder Nichtbeachtung der Betriebsanleitung entste hen Insbesondere kann der Hersteller nur dann f r...

Страница 12: ...g to translation of this documentation by Mahr GmbH The following symbols are used in these operating instructions General information Important information Failure to follow instructions marked with...

Страница 13: ...after March 23 2006 can be returned to us We will dispose of this equipment in an environmentally friendly way in accordance with the applicable EU Directives WEEE Waste Electrical and Electronic Equ...

Страница 14: ...ust be carried out by qualified personnel only 5 Any modification or manipulation of the amplifier requires the express written ap proval of Mahr GmbH and must be carried out by qualified personnel If...

Страница 15: ...ating instructions Contents Permitted Uses 1 Disposal 2 Voided Warranty 2 Safety Instructions 3 1 Introduction 6 2 Operating Elements 6 3 Working with 1901 TA 7 3 1 Zero adjustment 7 3 2 Adjusting the...

Страница 16: ...ut must be larger than 5 kOhm whereas the load across the current output must not exceed 2 kOhm The voltage output may be used for cable lengths of up to 10 m Please note that for cable length above 2...

Страница 17: ...lways be fixed mechanically such that the probe signal is about zero In addition the carrier frequency amplifier may be adjusted tuned to this zero point The zero setter features a setting range of ap...

Страница 18: ...the potentiom eter marked with Verst rkung Amplification in Fig 3 Two calibrated gage blocks are required for readjustment Readjusting the amplification is done on the open instrument with the power...

Страница 19: ...e open but inactive instru ment Use tweezers for relocating the jumpers Changing the measuring range Measuring range BR 1 BR2 125 m 1 2 1 2 250 m 3 4 1 2 500 m 5 6 1 2 1 000 m 7 8 1 2 2 000 m 9 10 1 2...

Страница 20: ...125 x 43 x100 mm with rack fixing 170 x 43 x 100 mm Distance of fixing bores 149 mm Fixing screws M5 head 10 mm Measuring ranges 125 m adjustable 250 m 500 m 1 000 m 2 000 m Output voltage 5V adjusta...

Страница 21: ...warranty does not cover natural wear and tear or defects caused by incorrect handling improper use or failure to comply with the operating instruc tions In particular the manufacturer shall only be he...