Mahr GmbH, MarSurf M 300 and RD 18
14
3
Commissioning
The MarSurf M 300 is delivered with the power
supply isolated to ensure that the builtin battery
does not lose its charge if stored for a lengthy
period.
1. Move the reset switch on the MarSurf
M 300
(12)
and the RD 18
(29)
to the right
into the "I" position to make both instru
ments ready for operation.
3.1
Switching on
the Meas uring Instruments
For a Bluetooth connection:
Please follow the instructions for set
ting up a Bluetooth connection given in
Section 4.3.3.
1. Hold down the green key
(23)
of the
MarSurf RD 18 approx. 1 second.
The LED next to the Bluetooth symbol
(22)
lights up in blue and indicates operational
readiness.
2. Then hold down the green
START
key of
the MarSurf M 300 approx. 1 second.
3. The MarSurf M 300 creates a Bluetooth
connection with the most recently selected
RD 18. While the connection is being estab
lished, this symbol is displayed:
When the connection has been created, the
LED below the Bluetooth symbol (1) lights
up and the
symbol is displayed in the
top status bar.
i
If a cable connection already exists:
1. Hold down the green
START
key of the
MarSurf M 300 approx. 1 second.
The LED next to the Bluetooth symbol
(22)
of the MarSurf RD 18 lights up in blue for a
short period and indicates operational readi
ness.
3.2
Switching off
the Meas uring Instruments
To switch the MarSurf M 300 and RD 18 off (if
they do not turn off automatically
10
):
1. Hold down the
START
key of the MarSurf
M 300 and press the
M
key.
Or hold down the green key
(23)
on the
MarSurf RD 18 approx. 3 seconds.
The MarSurf M 300 and the selected
MarSurf RD 18 are switched off.
2. To switch off other drive units:
Hold down the respective green key
(23)
on
the MarSurf RD 18 approx. 3 seconds.
3. To move the drive unit and pickup to the
safe parking position for transport:
Push the drive unit forward out of the hous
ing, turn it 180° to the right (stylus tip points
upwards) and pull it back into the housing
(guide pin engages with guide groove) un
til the rear lockin position is reached
(see
Fig. 14 in Section 6.2)
.
To switch off the instruments completely
(disconnect them from power), move
the reset switches
(12)
and
(29)
to the
left into the "0" position.
The instrument settings, including date
and time, are retained.
10 The instruments do not automatically turn off if the tim
eout is switched off (i. e. if the "Timeout" check box is not
activated in the "Basic settings" menu) or if the MarSurf
M 300 is connected to the mains with the AC adapter.
i