67
Mahr GmbH • Millimar C1202
The display shows information on
how to position the probe for the
second measuring point.
–
Use the
4
key to confirm the value or
Ó
to cancel the process.
–
Following successful measurement,
press
4
to activate the customer
calibration or
Ó
to cancel the
process.
An error message appears if the
measured values of the customer
calibration are not plausible.
– Use
4
to repeat the customer
calibration or
Ó
to cancel the
process.
To avoid any measurement errors that may be caused by a previously created
customer calibration, it is advisable to activate the factory standard calibration
again. This is always stored in the measuring instrument and is not changed by the
customer calibrations.