58
Mahr GmbH • Millimar C1202
5.2.6 Start delay
–
Select the
Start delay
submenu
using the
▼
or
▲
key.
–
Select using
4
key or exit submenu
using the
Ó
key.
–
Use the
▼
key to select the digit and
the
▼
or
▲
key to enter the relevant
value.
–
Use the
4
key to confirm the value or
Ó
to exit the submenu.
If you want a certain amount of
time to pass between the moment
(Start) is pressed and the
measurement starts, you can enter
a time period by which the start
of the measurement is delayed.
A start delay is particularly useful
when a dynamic measurement (e.g.
on a shaft) is performed and it is
supposed to start automatically
via a switch contact on the
measuring device. When the shaft
is approached, vibrations occur in
the measuring device that affect the
measuring result. During the delay
time, the vibrations subside and no
longer affect the measuring result.