Magtrol Load Monitoring Units LMU 212 & 217
Chapter 2 – Installation / Configuration
1. Place the jumper according to the selected supply mode (see
figure 2–5
) :
- 'DC' for a DC type supply with a voltage between 20 and 32 VDC
- 'AC' for an AC type supply (230 VAC, 115 VAC or 48 VAC)
2. In case of an AC supply position the switch on (see
figure 2–5
) :
- '230 V' for an AC voltage of 230 VAC
- '115 V' for an AC voltage of 115 VAC
- for operating voltages of 48 VAC or 20 to 32 VDC the position of the switch is irrelevant.
Fig. 2–5 Jumper and switch configuration
3. Check the rating of the fuse inserted in its support according to the following table:
operating
voltage
supply terminals
fuse
switch
0 v (18)
n (19)
p (20)
¶
230 VAC
earth
neutral
phase
80 mAT
230 V
115 VAC
earth
neutral
phase
160 mAT
115 V
48 VAC
earth
neutral
phase
400 mAT
irrelevant
20 - 32 VDC
0 V
—
20 - 32 V
400 mAT
irrelevant
A fuse of each rating is supplied with each load monitoring unit.
WARNING! foR sAfeTy ReAsoNs IT Is ImpoRTANT To secuRe
The sTAbIlITy of The supply uNIT used ANd
RespecT The opeRATING volTAGe selecTed oN
The lmu.
note :
Do not forget to report the designation of the collected external
signals connected to the supply terminal on the configuration and
calibration form (see
Appendix A
).
Operating voltage
20 to 32 VDC
Operating voltage
48 VAC
Operating voltage
115 VAC
Operating voltage
230 VAC