!
Since solvents such as ether, alcohol are highly
flammable, they must be handled carefully. Be
sure to keep these chemicals away from open
flames or potential sources of electrical sparks -
for example, electrical equipment that is being
switched off. Also remember to always use
these chemicals only in a well-ventilated room.
2.
When cleaning parts other than the lenses, do not
use the above chemicals. If they are seriously dirty,
wipe with gauze moistened slightly with a solution of
neutral detergent.
3.
Never attempt to remove parts other than those
specified from the microscope for cleaning.
4.
When the microscope is not used, ensure that the
light source lamp is cool, cover the microscope with
a dust cover or put it in a wood case, and store in a
dry place.
5.
To clean the condenser, fully loosen the securing
knob , then remove the condenser by lowering it
with the condenser up-down movement knob , and
clean the upper lens of the condenser.
The lens can be cleaned using the same method as
described in Fig.3. Wipe it as lightly as possible.
Reverse the removal procedure to attach the con-
denser.
Fig.3
1
2
1
Содержание Theia fi
Страница 12: ...Power Cord Socket...
Страница 34: ...NOTES...
Страница 35: ...NOTES...
Страница 36: ......