DIAGNOSTICS MENU cont.
Extended Errors cont.
BIT 1: A one (1) will appear if the Series 801 fails the inter-
nal oscillator test. The remote probe electronics may
be defective. Consult factory.
BIT 2: A one (1) will appear if the Series 801 fails the open
probe test. Ensure proper connection of the white
wire between terminal block TB1 and the probe
screw terminal in the probe housing. Consult factory
if unsuccessful.
BIT 3: A one (1) will appear in this position if the Series 801
fails the shorted probe test. Remove the white wire
from the probe screw and measure resistance
between probe screw and ground.
BIT 4: A one (1) will appear in this position if the signal
between the probe electronics and transmitter is out
of range. A direct “short” across the probe will also
trigger Bit 4. Ensure proper connection and polarity
of wiring between probe housing and transmitter.
Press ENT.
Extended Errors
will be displayed. Press the
DOWN arrow to display Previous Menu. Ent will display the
current menu, Diagnostics. Press the DOWN arrow to dis-
play LapTop.
LAPTOP MENU
This menu is used to enter the start and end day range for
downloading of stored data log for the Series 801.
Measured Values
System Config.
I/O Config
Calibration
Diagnostics
Laptop
STRT
END
With
LapTop
displayed, press ENT to display
STRT
___ .
Enter a number of the day of data to be downloaded. Press
ENT and
END
will appear. Enter an ending day. Press ENT
and
Laptop
will be displayed. If the laptop is connected for
RS-232 and the baud rate is compatible as set in the
I/O
Config
, the unit will begin to download the start /end day
range. Please note: this is an updating 30 day format. It will
download the minimum, maximum (at times of occurrence),
and total flow data for selected days.
Refer to Page 5 for Wiring information. Any P.C. communi-
cation program may be used to capture this data. Start the
communications program. Start an ASCII download and fol-
low the above procedure. Refer to your communications
program documentation if needed.
DIAGNOSTICS MENU
This menu is used to provide the user with testing the func-
tionality and providing useful information for troubleshooting
of the unit.
Measured Values
System Config.
I/O Config
Advanced Configuration
Diagnostics
Test Relays
Test 4–20 mA Loop
Signal Values
View Version#
Extended Errors
Previous Menu
Test Relays
This menu provides a controlled contact closure for each
relay to test its integrity. A password is required.
Test 4–20 mA Loop
Once the desired output loop current value is entered on the
display, the Series 801 will drive the given loop current to an
external device for calibration purposes. A password is
required.
Signal Values
Used for advanced factory troubleshooting.
View Version
This is the version number of the software installed in the
801.
Extended Errors
Provides further information for the factory personnel in
troubleshooting.
An eight digit code will be displayed by pressing ENT at the
Extended Errors prompt. By reading this code from right to
left, information on both probe circuits can be obtained. The
first four bits (starting at the right) are associated with probe
circuit 1, and the second set of four bits is associated with
probe circuit 2, as shown below.
0 0 0 0
0 0 0 0
Probe 2
Probe 1
All zeros will appear on a properly functioning Series 801. A
one (1) will appear when an error is detected.
BIT
4 3 2 1
0 0 0 0
Internal Oscillator Test Failure
Open Probe Test Failure
Shorted Probe Test Failure
Signal Loss
Series 801
UNIT CONFIGURATION cont.
Series 801
23