background image

10

57-656 SIL Safety Manual for ECLIPSE Model 706

Step

Action

Use the DIAGNOSTICS menu to observe the present Echo Curve. Confirm that the ECHO Waveform is
normal. The echo curve is dependent on the type of probe, the installation conditions and the level of
process on the probe. Comparison of the present Echo Curve to the one stored at the time of commis-
sioning the unit gives additional confidence of the normal operation of the unit. Use of the DTM and
digital communications is necessary for comparison of echo curves. 

a.) Select DIAGNOSTICS/ ECHO CURVE/ View Echo Curve

i.) Observe the present Echo Curve, identify the characteristic portions of the waveform related

to the FIDUCIAL, Process level, End of Probe and other features.

ii.) Confirm that the FIDUCIAL appears acceptable. Confirm the FIDUCIAL is located where

expected.

iii.) Confirm that the signal from the process level appears normal and is located as expected. 
iv.) Verify that the baseline of the waveform is smooth and flat.

v.) Compare to Echo Curve from commissioning in the FIDUCIAL area. 

b.) Access the Fiducial Ticks and Fiducial Strength values in the menu: DIAGNOSTICS /

ADVANCES DIAGNOSTICS / INTERNAL VALUES

i.) Observe and record:

1.) Fiducial Ticks _____________
2.) Fiducial Strength _____________

ii.) Confirm that these values match the previous values.

1.) Fiducial Ticks differs within ±100
2.) Fiducial Strength differs within ±15

6

Perform two-point calibration check of the transmitter by applying level to two points on the probe and
compare the transmitter display reading and the current level value to a known reference measurement. 

7

If the calibration is correct the proof test is complete. Proceed to step 9.

8

If the calibration is incorrect, remove the transmitter and probe from the process. Inspect the probe for
buildup or clogging. Clean the probe, if necessary. Perform a bench calibration check by shorting the
probe at two points. Measure the level from the bottom of the probe to the two points and compare to
the transmitter display and current level readings. 

a.) If the calibration is off by more than 2%, contact the factory for assistance. 

b.) If the calibration is correct, the proof test in complete.

c.) Re-install the probe and transmitter. 

9

Restore loop to full operation. 

10

Remove the bypass from the safety PLC to restore normal operation.

Содержание Eclipse 706GWR

Страница 1: ...rature and pres sure It can be used in liquids slurries or solids with a dielectric constant in the range 1 4 100 to meet the safety system requirements of IEC 61508 Edition 2 0 2010 and IEC 61511 1 B...

Страница 2: ...nstructions 5 6 1 Systematic Limitations 5 6 1 1 Application 5 6 1 2 Environmental 6 6 1 2 1 Storage 6 6 2 Installation 6 6 3 Skill Level of Personnel 6 6 4 Necessary Tools 6 6 5 Configuration 7 6 5 1...

Страница 3: ...is reflected The transit time of the pulse is then measured via ultra high speed timing circuitry that provides an accurate measure of the liquid level The amplitude of the reflection depends on the...

Страница 4: ...is reading an incorrect level within the 4 20 mA range 2 deviation MAGNETROL defines a safe failure as one in which the 4 20 mA current is driven out of range i e less than 3 8 mA or greater than 21 5...

Страница 5: ...f the transmitter One of the following Electronic Device Description Files is also required if HART is used Manufacturer Code 0x56 Model 706 Device ID 0x56E0 device revision 1 DD revision 2 For device...

Страница 6: ...evice Offers configuration recommendations NOTE This SIL evaluation has assumed that the customer will be able to acknowledge an over or under current condition via the Logic Solver 6 3 Skill Level of...

Страница 7: ...for SIL applications Consult factory prior to making any changes LOOP CONTROL MODE Ensure this is set to ENABLED THRESHOLD Set to FIXED if used in a hydrocarbon application with any possibility of wa...

Страница 8: ...SIL Proof Test Firmware can only be upgraded by factory personnel 7 0 Recurrent Function Tests 7 1 Proof Testing 7 1 1 Introduction Following is the procedure utilized to detect Dangerous Undetected...

Страница 9: ...at the time of the proof test event times are calculated a Choose the menu DIAGNOSTICS Present Status i Present Status should indicate OK b Choose the menu DIAGNOSTICS EVENT HISTORY Event Log i Any FA...

Страница 10: ...ess the Fiducial Ticks and Fiducial Strength values in the menu DIAGNOSTICS ADVANCES DIAGNOSTICS INTERNAL VALUES i Observe and record 1 Fiducial Ticks _____________ 2 Fiducial Strength _____________ i...

Страница 11: ...declares as the manufacturer that the level transmitter Guided Wave Radar 4 20 mA Model 706 51x xxx is suit able for the use in safety instrumented systems according to IEC 61511 1 if the safety inst...

Страница 12: ...12 57 656 SIL Safety Manual for ECLIPSE Model 706 8 2 FMEDA Report exida Management Summary...

Страница 13: ...13 57 656 SIL Safety Manual for ECLIPSE Model 706...

Страница 14: ...the PFDAVG value needs to be 10 3 and 10 2 This means that for a SIL 2 application the PFDAVG for a 1 year Proof Test Interval of the Model 706 511x xxx is approximately equal to 6 7 of the range Thes...

Страница 15: ...endent on the subsystem itself and its operating conditions The assumption of a constant failure rate is based on the bathtub curve Therefore it is obvious that the PFDAVG cal culation is only valid f...

Страница 16: ...1 IEC 61511 Mod Guideline on the Implementation of ANSI ISA 84 00 01 2004 Service Policy Owners of MAGNETROL controls may request the return of a control or any part of a control for complete rebuildi...

Отзывы: