10
57-645 Eclipse Guided Wave Radar Transmitter - PROFIBUS PA
™
2.4
QuickStart Configuration
The Eclipse transmitter comes partially configured from
the factory but can be reconfigured in the shop (disregard
fault message due to unattached probe). The minimum
configuration instructions required in the field follow. Use
the information from the operating parameters table in
Section 2.1.2 before beginning configuration.
1. Power up the transmitter.
The display changes every 5 seconds to show one of two
values: Status & Level.
2. Remove the cover of the lower electronic compartment.
3. Use the Up or Down Arrow ( ) keys to move from one
step of the configuration program to the next step.
4. Press the Enter Arrow ( ) key. The last
character in the first line of the display
changes to an exclamation point (!).
5. Use the Up or Down Arrow ( ) keys to increase or
decrease the value in the display or to scroll through the
choices.
6. Press the Enter Arrow ( ) key to accept a value and move
to the next step of the configuration program (the default
password is 0).
7. After entering the last value, allow 10 seconds before
removing power from the transmitter.
The following configuration entries are the minimum required for
configuration (the default password is 0 from the LCD/keypad).
PrbModel
xxx
PrbModel
(select)
Probe Ln
xxx.x
LvlOfst
xxx.x
Dielctrc
(select)
Select the
Probe Model
to be used
Model 705:
7xA-x, 7xB-x, 7xD-x, 7xE-x, 7xF-F, 7xF-P,
7xF-4, 7xF-x, 7xJ-x, 7xK-x, 7xP-x, 7xQ-x, 7xR-x,
7xS-x, 7xT-x, 7x1-x, 7x2-x, 7x5-x, 7x7-x
Select the type of
Probe Mounting
to vessel (NPT, BSP, or
flange).
Select from Level Only, Level and Volume, Interface Level
or Interface Level and Volume.
Enter the exact
Probe Length
as printed on the probe
nameplate.
Enter the
Level Offset
value. Refer to Section 2.6.6 for
further information. (The unit is shipped from the factory
with offset = 0; i.e., all measurements are referenced to
the bottom of the probe).
Enter the
Dielectric
range for the material to be measured.
Level Offset
Probe Length
Probe Mount
Probe Model
Dielectric
of Medium
2
5
6
4
1
➪
➪
➪
➪
➪
➪
①
②
③
④
⑤
⑥
PrbMount
(select)
Enter
Down
Up
MeasType