1/3
LH71 Initial setup (Example)
Ver.1 (2021.1) CS&S
Press key once.
The display will be . (Measurement unit resolution)
Press key to check the current measurement unit resolution.
If the resolution is not 5μm, press key to change it.
After checking the resolution, press key to save the settings.
Switch from basic settings to normal count mode.
Press
key once and the display will change to
.
Press
key once and the display will change to
.
Press
key once to automatically switch to advanced settings mode.
The display will be
. Press
key once.
Confirm that the display is
display.
(Press
key to change the polarity)
When the selection is complete, press
key to save the settings.
When the display is , press the key on the right side
once.
The display will be . (Destination selection mode)
( If you want to change the destination, press key. )
When you have made your selection, press to save your settings.
Setting conditions
Connect the AC adapter and
Turn on the primary power supply
(Not key)
Confirm that the resolution of input 1
is 5 μm (SJ700A scale).
* 10 μm “10U” for Digiruler PL20C
When you press
key, the display becomes the normal display.
This completes the initial settings
Input 1
For LH71-1, LH71-2
LH71 Initial setup (Example)
AC adapter is energized.
Turn on
key,
the LED light changes
from “blinking” to “off” and then
Press and hold
key
for 3 seconds.
The following display appears
Ver.1 (2021.1) CS&S
Setting items
Setting value (selection)
Basic
settings
TYPE
1 axis model is not displayed
ADD(Addition display)
1 axis model is not displayed
Destination
STD
Measuring unit
resolution
5μ
m
Detailed
settings
Display resolution*
and polarity
5μ
m
Display switching
Input 1:
Axis label
Display
:
X
Scaling
1 time
(
1.000000)
Compensation value
No setting
(
OFF
)
Flicker control
Strong
(
2
)
Sleep
No setting
(
OFF
)
Measuring unit
SJ700A (1 axis)
Output resolution: 5μm
When the power is turned
on for the first time after
shipment from the factory
When you want to change
the basic settings
START
Set to basic setting mode
The detailed settings are the factory default settings.
Display unit: LH71-1 (1 axis input)
SJ700A
X axis
Finished
* The initial value is the measurement unit resolution set in the basic settings.
default
default
default
default
default
default
AC primary power source
①
④
③
②
default
About 2 sec
Power ON