6. Check UV output of lamp using Magnaflux black light meter. Measure intensity 15" from the lens; clean
lens or replace bulb if reading is less than 1000µW/cm².
WARNING:
Lens on the black light may be
extremely hot. Let the unit cool before cleaning or replacing bulb.
7. Set Pressure Regulator Valve to 40 psig.
WARNING: Lowering air pressure setting may be
required for small part to prevent damage.
8. If equipped with Auto-bath (Optional), toggle the Auto-bath button, and the PUMP button to off. If
equipped with the Auto-Mag, toggle Auto-Mag to off by pressing the Auto-Mag button, then releasing.
9. Check operation of headstock clamping cylinder by pressing and releasing Foot Switch or Optional
Dual Palm.
10. Adjust Speed Control Needle Valve on headstock for desired clamping speed.
11. Press CONTACT push-button.
12. Set the amperage using the CONTACT current control or use the AMPERAGE SETPOINT on the HMI
to a desired value.
13. Check for magnetizing current between contacts.
a. Place a suitable part between headstock and tailstock contacts.
b. Place a piece of insulating material, such as a thick piece of cardboard, between headstock contact
and the end of the part.
c. Press and release Foot Switch or Optional Dual Palm to clamp part.
d. Press MAG push-button while observing Digital Ammeter.
- The current assurance light should not light.
f. If there is any indication of current on the ammeter, an electrical short exists. For corrective action,
refer to Section 5 for Troubleshooting.
g. Place a minimum 1" diameter test bar between headstock and tailstock contacts.
h. Press and release foot switch or Optional Dual Palm to clamp test bar.
i. Press MAG Current push-button while observing ammeter.
- The magnetizing current flowing through the test bar should appear on the Digital Ammeter.
- The Green Current Assurance Light on the Ammeter (HMI) should light indicating current flow
through the test bar.
- For MOD AC (Optional), repeat Step 13.i with AC selected on the HMI.
14. Press and release foot switch or Optional Dual Palm to release the part.
15. If equipped with Auto-bath option, check as follows:
a. Press Auto-bath button and PUMP button to ON.
b. If equipped with foot switch, follow steps 13 and 14, keeping in mind that the foot switch is pressed,
and released to clamp the part, then the bath applicator will extend to apply bath. When extend limit
is reached, it will retract and the bath will turn off, and the mag shot will occur.
c. Part will then unclamp signifying that the Auto-bath cycle is complete.
d. If equipped with Optional Dual Palm, follow steps 13 and 14, keeping in mind that the Dual Palm
are pressed to clamp the part and extend the bath applicator. After mag shot occurs, headstock
Содержание D-2000 Series
Страница 1: ...D 2000 Series Revised March 2020 Page 1 0 D 2000 SERIES ...
Страница 9: ...D 2000 Series Revised March 2020 Page 1 8 PAGE LEFT BLANK INTENTIONALLY ...
Страница 20: ...D 2000 Series Revised March 2020 Page 2 11 PAGE LEFT BLANK INTENTIONALLY ...
Страница 23: ...D 2000 Series Revised March 2020 Page 3 3 Figure 3 4 Figure 3 5 10 11 12 14 13 1 17 2 18 19 15 16 20 23 21 22 ...
Страница 33: ...D 2000 Series Revised March 2020 Page 3 13 PAGE LEFT BLANK INTENTIONALLY ...
Страница 40: ...D 2000 Series Revised March 2020 Page 4 7 PAGE LEFT BLANK INTENTIONALLY ...
Страница 45: ...D 2000 Series Revised March 2020 Page 5 5 PAGE LEFT BLANK INTENTIONALLY ...
Страница 48: ...D 2000 Series Revised March 2020 Page 6 3 PAGE LEFT BLANK INTENTIONALLY ...