![MagListo 5M Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/maglisto/5m/5m_user-manual_3511927012.webp)
9
www.bioneer.com
BQ-042-101-05
Revision : 2 (2016-07-04)
18.
Attach the magnet plate to
MagListo
TM
rack and invert the tube 3 ~ 4 times gently until the beads
tightly bind to the magnet.
19.
Without removing the tube from
MagListo
TM
rack, carefully take the supernatant containing DNA to a
sterile microcentrifuge tube.
20.
Discard the used magnetic nano beads. Do not reuse the beads.
Summary of Reagents Volume in Each Step of DNA Extraction from Whole Blood
Step
Buffer
Mini
Midi
Maxi
Page
Blood (+ PBS)
200 μl
2 ml
4 ml
P. 6
Lysis
Buffer ② (Binding)
200 μl
2 ml
4 ml
P. 6
Precipitation
Absolute Ethanol
400 μl
4 ml
8 ml
P. 6
Bead Binding
Magnetic Nano Bead - DNA
100 μl
500 μl
1 ml
P. 6
1
st
Washing
(Repeat)
Buffer ③ (1
st
Washing)
500 μl
3 ml
6 ml
P. 7
2
nd
Washing
Buffer ④ (2
nd
Washing)
700 μl
5 ml
10 ml
P. 8
3
rd
Washing
Absolute Ethanol
700 μl
5 ml
10 ml
P. 8
Elution
Buffer ⑤ (Elution)
100 μl
500 μl
1 ml
P. 8