About Calibration
Why Calibration is Important
The analyzer must be calibrated to the lot of sensors to ensure accurate results.
You must calibrate the analyzer:
•
The first time you use the analyzer.
•
Each time you start a new lot of test kits.
•
Any time the analyzer displays a recalibration message.
The Calibration Button
Each LeadCare II Test Kit comes with 48 sensors and a calibration button. The
button transfers calibration and expiration information to the analyzer. When you
touch the calibration button to the button reader, you will hear an audible beep.
The lot number appears on the screen to verify that the button was read properly.
The CALIBRATION SUCCESSFUL message flashes for 2 seconds.
Calibrating the Analyzer
CAUTION:
•
Calibration is required for each new lot of test kits. Use only the
calibration button packaged with the test kit you are using.
Failure to use the correct calibration button could cause
inaccurate results.
•
Do NOT use items from more than one test kit at a time.
•
Each test kit comes with a calibration button marked with the
sensor lot calibration code. Always make sure that the lot
numbers on the sensor container and calibration button match
the SENSOR LOT number on the analyzer display.
See calibration instructions below.
1. Turn on the analyzer. The
analyzer is ready when the
“Prepare Sample” message
appears.
NOTE:
The first time you turn on
the analyzer, you will see the
“PLEASE CALIBRATE” message
.
PREPARE SAMPLE
USE SENSOR LOT XXXX
OR RECALIBRATE
THEN INSERT SENSOR
Содержание LeadCare II
Страница 8: ...vi This page intentionally left blank...
Страница 24: ...LeadCare II User s Guide 1 16 This page intentionally left blank...
Страница 30: ...LeadCare II User s Guide 2 6 This page intentionally left blank...
Страница 36: ...LeadCare II User s Guide 3 6 This page intentionally left blank...
Страница 56: ...LeadCare II User s Guide 6 2 This page intentionally left blank...
Страница 60: ...LeadCare II User s Guide A 4 This page intentionally left blank...
Страница 64: ...LeadCare II User s Guide B 2 This page intentionally left blank...
Страница 65: ...LeadCare II User s Guide C 1 Appendix C...
Страница 66: ...LeadCare II User s Guide C 2 This page intentionally left blank...
Страница 68: ...LeadCare II User s Guide D 2 This page intentionally left blank...
Страница 70: ...LeadCare II User s Guide E 2 This page intentionally left blank...