Chapter 2 Calibrating the Analyzer
This chapter describes how to calibrate the analyzer. The analyzer must be
calibrated to the lot of sensors to ensure accurate results. This chapter contains
the following topics:
•
Turning the Analyzer On and Off
•
Analyzer Self-Tests
•
About Calibration
•
Calibrating the Analyzer
CAUTION:
•
Calibration is required for each new lot of test kits. Use the
calibration button in the test kit. Use only the button
packaged with the kit you are using. Failure to use the
correct calibration button with the test kit could cause
inaccurate results.
•
Do NOT use items from more than one test kit at a time.
•
Always make sure that the lot numbers on the sensor
container and calibration button match the SENSOR LOT
number on the analyzer display
.
Содержание LeadCare II
Страница 8: ...vi This page intentionally left blank...
Страница 24: ...LeadCare II User s Guide 1 16 This page intentionally left blank...
Страница 30: ...LeadCare II User s Guide 2 6 This page intentionally left blank...
Страница 36: ...LeadCare II User s Guide 3 6 This page intentionally left blank...
Страница 56: ...LeadCare II User s Guide 6 2 This page intentionally left blank...
Страница 60: ...LeadCare II User s Guide A 4 This page intentionally left blank...
Страница 64: ...LeadCare II User s Guide B 2 This page intentionally left blank...
Страница 65: ...LeadCare II User s Guide C 1 Appendix C...
Страница 66: ...LeadCare II User s Guide C 2 This page intentionally left blank...
Страница 68: ...LeadCare II User s Guide D 2 This page intentionally left blank...
Страница 70: ...LeadCare II User s Guide E 2 This page intentionally left blank...