 Product User Guide  |  6

UltraShock

DEVICE OPERATION

Logging Mode - Peak vs. Instantaneous 

When 

Custom Start

 is selected in the MadgeTech 4 Software, the user is presented with a range of logging time intervals or 

actual recording rates (sampling rates). For all but the two highest recording rates, the user will also have a choice of Peak 
vs. Instantaneous capture.

PEAK

INSTANTANEOUS

Regardless of the logging interval, the accelerometers will 
internally sample at 512Hz. At the logging interval (for example, 
30 seconds), the Peak value found in the previous logging 
interval is stored.

    

The red markers show the peak value for the interval which will 
be stored.

The accelerometers will only sample data at the logging rate. For 
instance, if a 30 second logging rate was selected, then only a 
single data point is captured at each 30 second instant.

     
The potential to lose valuable data can be seen. This, of course, 
is very dependent on the applied shock duration and wave 
shape. It’s possible for a significant shock event to occur and not 
even be visible in the data for the Instantaneous mode.

1.  Connect the logger to the interface cable.
2.  Highlight the data logger in the Connected Devices list. 
3.  Click 

Custom Start

.

4.  Choose the 

Start method

—Now or Delay (select date 

and time).

5.  Choose the 

Stop method

—Manual or Automatic (select 

date and time).

6.  Select the 

Reading interval

.

7.  Select the 

Shock recording mode

 (if applicable)—

Instantaneous or Peak

NOTE:

  When switching USB connections from a PC, laptop or power adapter, users can expect a 2 minute delay.

Содержание UltraShock

Страница 1: ...ck Data Logger PRODUCT USER GUIDE TABLE OF CONTENTS 2 Product Overview 4 Software Installation Guide 5 Device Operation 8 Product Maintenance 9 Specifications 10 Dimensional Drawings To view the full...

Страница 2: ...occur when the UltraShock is mounted by two bolts directly to a solid metal flat surfaced object of significant mass For various applications the rigid bolted down mounting may not be optimal or prac...

Страница 3: ...SB connected and adequate battery charge level Flashes every 15 seconds maximum Indicates low battery charge level Logging Data with USB communication Flashes every 15 seconds Indicates normal operati...

Страница 4: ...ts before other parameters Pressure The UltraShock pressure sensor is designed to measure barometric pressure from very low altitudes to those altitudes found in aircraft up to high altitude balloons...

Страница 5: ...pending upon your start method 6 Disconnect the data logger from the interface cable and place it in the environment to measure Note The device will stop recording data when the end of memory is reach...

Страница 6: ...s will only sample data at the logging rate For instance if a 30 second logging rate was selected then only a single data point is captured at each 30 second instant The potential to lose valuable dat...

Страница 7: ...100 samples are acquired at the selected logging rate The trigger sample count should be chosen carefully because it may gather data points for which we have no interest For instance if 100 were enter...

Страница 8: ...cell pack Please note the safe handling and storage recommendations in the specifications section Inspection The UltraShock may be inspected visually for any signs of damage especially to the gray gas...

Страница 9: ...nto the device to restrict access to configuration options Data may be read out with the password Calibration Digital calibration through software Calibration Date Automatically recorded within device...

Страница 10: ...21 8 mm 1 72 43 7 mm 0 20 5 1 mm 2X 0 20 5 1 mm THRU 4X R0 25 R6 4 mm APPLIQUE COVER GASKET BATTERY PACK RH AND PRESSURE SENSOR PORT TYPE C USB CONNECTOR 3 00 76 2 mm RED LED INDICATOR GREEN LED INDIC...

Отзывы: