MadgeTech, Inc.
PO Box 50
l
Warner, NH 03278
Phone 603.456.2011
l
Fax 603.456.2012
www.madgetech.com
l
Red 485
Dark Blue Pantone 289
Light Blue 58% of Pantone 289
REV A 2008.07.08
To view the full MadgeTech product line,
visit our website at
www.madgetech.com.
Red 485
Dark Blue Pantone 289
Light Blue 58% of Pantone 289
Product Information Card
RFRHTemp101A
Product Quick Reference Card
RFRHTemp101A
Temperature and Humidity Recorder and Wireless Transmitter
Specifications subject to change.
See MadgeTech’s terms and conditions at www.madgetech.com
Part Number
RFRHTemp101A
Internal Temperature Sensor
Internal Semiconductor
Internal Temperature Range
-30 to +70°C
Internal Temperature Resolution
0.1°C
Internal Temperature Accuracy
+0.5°C
RH Measurement Range
0 to 95%RH
RH Resolution
0.5%RH
RH Calibrated Accuracy
+3%RH
Memory
5,461/channel
Reading Rate
30 seconds up to 12 hours
LED
Red and Green
Required Interface Package
RFC101A
PC Baud Rate
57,600
RF Carrier Frequency
418 + 0.075MHz
RF Baud Rate
4,800
RF Range
Typical Outdoors/Line of Sight: Up to 120’ (36m)
Typical Indoors/Urban: Up to 40’ (12m)
Typical Battery Life
1 Year
Material
ABS Plastic
Operating Environment
-30 to +70°C, 0 to 95%RH (Non-Condensing)
Dimensions
1.7” x 2.3” x 0.8” plus 2.0” Antenna
(43mm x 59mm x 21mm plus 51mm Antenna)
Approvals
US (FCC), CA (IC)
Device Maintenance
Battery Replacement
Materials:
Small Phillips Head Screwdriver
Replacement Battery (LTC-7PN)
- Puncture the center of the back label with the screw driver
and unscrew the enclosure.
- Remove the battery by pulling it perpendicular to the circuit
board.
- Insert the new battery into the terminals and verify it is secure.
- Screw the enclosure back together securely.
Note: Be sure not to over tighten the screws or strip the threads.
Battery Warning
WARNING: FIRE, EXPLOSION, AND SEVERE BURN HAZARD. DO NOT
SHORT CIRCUIT, CHARGE, FORCE OVER DISCHARGE, DISASSEMBLE,
CRUSH, PENETRATE OR INCINERATE. BATTERY MAY LEAK OR
EXPLODE IF HEATED ABOVE 80°C (176°F).
Recalibration
The RFRHTemp101A standard calibration is one point at 25°C and
two points at 25%RH and 75%RH.
Pricing:
Recalibration traceable to NIST
$90.00
Recalibration
$70.00
Additional:
As Found Data
$15.00
Additional Points
$15.00 per point
To send the devices back, visit www.madgetech.com, select
Services
then
RMA Process
.