background image

1 www.macurco.com 

Rev: 2/06 

 

MACURCO GAS DETECTORS

 

 

DMK-1 

INSTALLATION 

AND OPERATION INSTRUCTIONS

 

WWW.MACURCO.COM

 

 

 
GENERAL INFORMATION 
 

 

The DMK-1 is a duct mount enclosure that provides the means to duct mount all of Macurco’s 

standard 4 x 4 mount gas detectors. By duct mounting a gas detector, the unit is able to monitor gas 
concentrations in ventilation ducts. 
 

THE FOLLOWING PARTS ARE INCLUDED WITH THE DMK-1: 
 

Enclosure 

with 

cover 

and 

gasket 

    1 

EMT Setscrew Adapter 

 

 

 

 

 

¼” X 

 

6” 

Exhaust 

Tube 

     1 

¼” X 1.1” hex aluminum 

spacers 

    4 

Machine Screws 4/40 x ½” phillips flat head  

 

 

Machine Screws 4/40 x 3/8” phillips pan head 

 

 

Strain relief / seal for wiring exit hole 

 

 

 

Nut 

for 

strain 

relief      1 

Sheet metal mounting screws #6 X 1” 

 

 

 

Adhesive 

foam 

compression 

seal 

    2 

full 

size 

drill 

template 

     1 

1/2” sampling tube (outer diameter) 

not provided

 

 

 

 

 

INSTALLATION 

 

1.  Open the enclosure by removing the 4 black integrated cover screws. 
2.  Install the setscrew adapters into the 2 (entrance and exit) holes drilled into the back of the enclosure 

so that the setscrew side of the adapter is on the outside of the enclosure. 

3.  Decide which hole will be used for sampling and which will be used for exhaust. The sampling tube 

(

not provided

) should be on the ”upstream” side of the enclosure, as it is to be mounted on the duct 

(“upstream” refers to the flow of air in the duct * see 

reference diagram

 below*).  

4.  Install the exhaust tube into one of the setscrew adapters so that the squared end is mated to the 

setscrew adapter and the pointed end is toward the duct. The tube should be turned so that the longer 
pointed side is facing the direction of  “upstream”  air flow. 

5.  Install the sampling tube (

not provided

) into the other setscrew adapter according to the sampling tube   

manufacturer’s instructions. 

6.  Install the detector onto the top of the 4 hex aluminum spacers. Secure it in place using the 4 pan head 

machine screws. Run the wiring through the strain relief. Twist the outer cap of the strain relief 
clockwise to compress the gasket material around the wiring. 

7.  Mark and drill the 2 large (1 ½”) and 4 small (3/32) holes in the duct you want to mount the unit on, 

using the template provided. The two large holes are for the sampling and exhaust tubes. The four 
small holes are starter holes for the sheet metal mounting screws. Ensure that the sampling and exhaust 
tube holes are large enough to allow the setscrew adapters to fit. 

Отзывы: