background image

 

IMPORTANT: Keep these user instructions for reference. 

AimSafety™ PM

400

 

Multi-Gas Personal Monitor 
User Instructions 

 

 

 

 

Содержание AimSafety PM400

Страница 1: ...IMPORTANT Keep these user instructions for reference AimSafety PM400 Multi Gas Personal Monitor User Instructions...

Страница 2: ......

Страница 3: ...Down 10 4 1 Charging the Device 10 4 2 Power On the Monitor 10 4 3 Powering Off the Monitor 11 4 4 User Interface 11 4 4 1 Display Modes 11 4 5 Menu Screens 12 4 6 Alarm and Alerts 14 4 6 1 Alarm and...

Страница 4: ...or sensor to malfunction Alarm values are set based on the alarm standard that are required by international standards Therefore alarm values should be changed only under the responsibility and approv...

Страница 5: ...gases The PM400 s audible visual and vibrating alarms alert the user of unsafe gas concentrations Standard sensors include Carbon Monoxide Hydrogen Sulfide Oxygen and either Catalytic Bead PM400 P or...

Страница 6: ...ver Range Alarm sounds exit to a safe area immediately A simple press of the Function key acknowledges and silences the alarm The sensor continues to stay active even though the level has exceeded 100...

Страница 7: ...hydrogen and or acetylene The PM400 IR does NOT detect these gases 1 3 3 2 Advantages of NDIR technology The NDIR sensor does not require oxygen to detect gases Because the NDIR sensor uses absorbed l...

Страница 8: ...Logging Data logging Two months Continuous at 1 minute intervals 24 hours a day 7 days a week Event logging 30 events Calibration logging 30 calibrations Bump Test logging 30 bump tests Mounting type...

Страница 9: ...PM400 Multi Gas Personal Monitor REV 2 0 34 2900 0206 2 8 P a g e 3 Product Overview 3 1 Monitor Overview 3 1 PM400 Monitor 3 2 Display Overview 3 2 PM400 Display Screen...

Страница 10: ...2 9 P a g e 3 2 1 LCD display symbols High Alarm Display Zero Calibration Display Low Alarm Display Device Stabilization Calibration Succeeded Alarm Condition Standard Gas Calibration Display STEL Ala...

Страница 11: ...to a standard wall socket The battery meter appears on the right side of the display and repeatedly cycles upward to the top and the LEDs will flash continuously NOTE If your device is set to Stealth...

Страница 12: ...ser Interface The PM400 has two operational modes Measuring Mode Standard display operation with real time gas readings always displayed Safe Zone Mode Displays Safe Zone unless one or more gas concen...

Страница 13: ...ess the Function key to advance to Stealth Mode if enabled Stealth Mode is added to the menu to indicate that the audible and vibrating alarms are disabled Display flags function normally Peak MIN Pre...

Страница 14: ...d hold the Function key for the seconds to enter the Alarm Values menu Press the Function key to advance the display through each of these in order Firmware Press the Function key to advance to the Fi...

Страница 15: ...oint the alarms will activate The display will show the appropriate alarm icon s High Low TWA STEL Over Range OVL and the gas level The monitor will vibrate the buzzer will sound and the LEDs will fla...

Страница 16: ...Icons and gas concentration TWA Alarm Exceeds TWA alarm set point Icons and gas concentration STEL Alarm Exceeds STEL alarm set point Icons and gas concentration Over Range Alarm Detected gas es exce...

Страница 17: ...changed using the PM Link and software or the Bump Test Calibration Station Gas Type Combustible LEL CO H2S O2 Low Alarm 10 LEL 35 PPM 10 PPM 19 5 High Alarm 20 LEL 100 PPM 20 PPM 23 5 TWA Alarm 35 PP...

Страница 18: ...if it fails a Bump Test To perform a Bump Test 1 On the Measuring screen make sure the oxygen sensor reads 20 9 and the other gases read zero 2 Clip the calibration cap to the unit The cap attaches to...

Страница 19: ...second countdown timer appears in the lower right corner while the rest of the display alternates between GAS IN and real time gas measurements 9 Watch the display as the timer counts down Results sho...

Страница 20: ...settings and re test the unit If everything looks correct and the unit fails perform a Zero calibration followed by a Span calibration Do not use the monitor until the reason for the discrepancy for t...

Страница 21: ...is performed on all devices prior to shipment Once received calibration should be performed monthly or quarterly depending on frequency of use All alarms are muted during calibration 6 1 Accessing th...

Страница 22: ...ten second countdown displays in the lower right corner of the display while the device cycles through displaying detected gas levels 4 Upon a successful calibration your device displays the Calibrati...

Страница 23: ...resh air calibration see section 6 2 Fresh Air Calibration 2 Clip the calibration cap to the unit The cap attaches to the recesses in either side of the unit You should hear a click when the cap is se...

Страница 24: ...the timer to reach zero 9 After a successful calibration the Calibration Success screen displays 10 After the calibration is complete be sure to turn off the calibration gas If calibration fails the...

Страница 25: ...18 0 Vol 100 ppm 25 ppm 50 LEL 6 4 Exiting the Calibration Menu To exit the Calibration menu 1 Press the Function key until ESC is displayed on the screen 2 Press and hold the Function key for 5 secon...

Страница 26: ...the event Sensor Readings Alarm Type Serial Number of the Device The Bump Log records the following Date Time Audible alarm Sensor Readings Alarm Type Alarm Pass Fail Serial Number of the Device The C...

Страница 27: ...h a dry cloth may generate a static charge and result in an explosion if located in a hazardous environment Occasionally clean the monitor with a soft cloth Do not use detergents or chemicals If neces...

Страница 28: ...ia IIC T4 Ga IP67 KRH 17 ATEX 0013 Directive 2014 34 EU KCS Ex ia IIC T4 KTL 16 KA2BO 0457 INMETRO Ex ia IIC T4 Ga BVC16 5919 Standards The electrical apparatus and any acceptable variations to it sp...

Страница 29: ...PM400 Multi Gas Personal Monitor REV 2 0 34 2900 0206 2 28 P a g e 11 Appendix A Table of Figures 3 1 PM400 Monitor 8 3 2 PM400 Display Screen 8 5 1 Calibration Hood 17 6 5 Calibration Hood 22...

Страница 30: ......

Страница 31: ......

Страница 32: ...ng out of or related to the use of said gas monitor Manufacturer or its agent s liability shall be limited to replacement or repair as set forth above Buyer s sole and exclusive remedies are return of...

Отзывы: