Manual

Instructions for use

  SBRT Systems

3

15112021V22IFU123100

MacroMedics ®

1.  Warnings

•  Inspect the device thoroughly before use. Do not use the device if damaged or if missing parts. 

•  Check individual components on MR-safety when used together in a patient positioning system. 

•  Do not use MR unsafe items in an MR environment.

•  Ensure that the baseplate is correctly secured to the couch prior to use.

•  Do not lean on the OmniBoard HXP Rail System.

•  If the cranial end of the board is extended past the couch, do not exceed a load of 15 kg in this 

area.  

•  Do not exceed an equally distributed load of 200 kg on the baseplate.

•  Take extra care when securing the baseplate to the couch to not damage the products.

•  Prevent the surface from damage and scratching. 

•  Rulers used on the devices are only intended for reference of the device positioning of accessories. 

Rulers shall not replace patient position verifi cation.

•  Keep the vacuum cushion valve out of the radiation treatment path.

• 

Do not infl ate the vacuum cushion above 100 Mbar.

•  Please be aware not to deform the cushion whilst the patient is being (re)positioned. 

•  Place the patient carefully on the device and make sure that they lie down calmly. Instruct the 

patient not to reposition themself after set-up.

•  Prevent the formation of closed-loops as they can lead to MRI burns. 

       Use an isolator between the patient’s arms to avoid skin-to-skin contact.

•  In the USA, federal law restricts this device to sale by or on the order of a licensed healthcare 

       practitioner.

2.  MRI Safety Information

2.1 

   MR CONDITIONAL

           

Non-clinical testing has demonstrated the following devices are MR conditional:

•      123450  Vacuum Cushion, abdom. S

•      123470  Vacuum Cushion, abdom. XL

•      125310  Rigid UAS MR set/2    

•      125350  Adjust. UAS MR set/2

• 

    130250  FeetSupport™, fi xed 

•      130260  FeetSupport™, rotatable 

•      130460  FeetSupport 

TM

, rotatable 2.0

•      130450  FeetSupport 

TM

, fi xed 2.0

The devices labelled as MR Conditional can be safely used in a MR system meeting the following condi-

tions:

• 

    Static magnetic fi eld of 1.5 Tesla and 3 Tesla, with 

• 

    Maximum spatial fi eld gradient of 2,500 G/cm (25 T/m) 

•      Maximum force product of 50,000,000 G²/cm (50 T

2

/m) 

Due to the size and location of the metallic parts, no clinically signifi cant RF heating is expected.

It is the responsibility of the user to ensure that these conditions are met. 

•      126890  OmniBoard™  2.0 base, EXF MRC

•      174550   OmniBoard™ base MRC

•      174990  OmniBoard™  2.0 S-Type MRC

•      175050  OmniBoard™ Base, S-Type MRC

•      175150  OmniBoard™ Base short S-Type MRC

•      175890  OmniBoard™ 2.0 S-Type Base MRC

Содержание SBRT

Страница 1: ...acroMedics BV Oostbaan 670 2841ML Moordrecht The Netherlands 31 0 182 389 777 info macromedics com www macromedics com MacroMedics SBRTTM Systems MANUAL 123100 Instructions for use English Instruction...

Страница 2: ...ontents 1 Warnings 3 2 MRI Safety Information 3 3 Intended Use 4 4 Installation Instructions All baseplates except HXP 5 5 Installation of Accessories 6 6 Installation Instructions HXP Baseplates 8 7...

Страница 3: ...d make sure that they lie down calmly Instruct the patient not to reposition themself after set up Prevent the formation of closed loops as they can lead to MRI burns Use an isolator between the patie...

Страница 4: ...BRT Secure Arch size S 123660 SBRT Secure Arch size S HXP 123700 SBRT Secure Arch size XS 123760 SBRT Secure Arch size XS HXP 123860 EAMIS Lite MR 3 Intended Use Positioning of the patient during radi...

Страница 5: ...ber MR safe 3 Pin CouchStrip II PLACE THE SBRT BASEPLATE Align the SBRT baseplate with the Couchstrips and slot it into place Make sure that no movement is possible between the baseplate and the couch...

Страница 6: ...ngravings on the SBRT Secure Arch for the cranial caudal direction III Rotate the knobs on each side of the SBRT Secure Arch until the clamps properly fasten onto the baseplate Do not apply too much f...

Страница 7: ...lock the Plate into position 5 3 PLACING THE ABDOMINAL VACUUM CUSHION I Slot the white pins of the cushion into the holes at the top of the SBRT Secure Arch and slide the white sliding mechanism into...

Страница 8: ...100 CouchStrips and Sidepieces II Placing the OmniBoard Align the Omniboard with the CouchStrips and slot it in place Make sure that no movement is possible between the baseplate and the couch 6 2 ATT...

Страница 9: ...ETACHING THE MULTIBOARD HXP RAIL SYSTEM The MultiBoard HXP Rail System can be attached to the MultiBoard HXP baseplate Align the bottom of the MultiBoard HXP baseplate with the rail system and slot it...

Страница 10: ...he device in a clean ventilated environment Dispose of the device in accordance with the local laws and regulations applicable to your country 8 Warranty The MacroMedics products are warranted to be f...

Страница 11: ...se Caution Indicates the need for the user to consult the instructions for use for important cautionary information such as warnings and precautions that cannot for a variety of reasons be presented o...

Страница 12: ...MacroMedics BV Oostbaan 670 2841ML Moordrecht The Netherlands 31 0 182 389 777 info macromedics com www macromedics com Visit www macromedics com for a digital copy of this manual...

Отзывы: