background image

MacroMedics BV

Oostbaan 670

2841ML Moordrecht

The Netherlands

+31(0)182 389 777

[email protected]

www.macromedics.com

Macro

Medics ®  SBRT

TM

  

Systems 

MANUAL 123100 

Instructions for use

English Instructions

Содержание SBRT

Страница 1: ...acroMedics BV Oostbaan 670 2841ML Moordrecht The Netherlands 31 0 182 389 777 info macromedics com www macromedics com MacroMedics SBRTTM Systems MANUAL 123100 Instructions for use English Instruction...

Страница 2: ...ontents 1 Warnings 3 2 MRI Safety Information 3 3 Intended Use 4 4 Installation Instructions All baseplates except HXP 5 5 Installation of Accessories 6 6 Installation Instructions HXP Baseplates 8 7...

Страница 3: ...d make sure that they lie down calmly Instruct the patient not to reposition themself after set up Prevent the formation of closed loops as they can lead to MRI burns Use an isolator between the patie...

Страница 4: ...BRT Secure Arch size S 123660 SBRT Secure Arch size S HXP 123700 SBRT Secure Arch size XS 123760 SBRT Secure Arch size XS HXP 123860 EAMIS Lite MR 3 Intended Use Positioning of the patient during radi...

Страница 5: ...ber MR safe 3 Pin CouchStrip II PLACE THE SBRT BASEPLATE Align the SBRT baseplate with the Couchstrips and slot it into place Make sure that no movement is possible between the baseplate and the couch...

Страница 6: ...ngravings on the SBRT Secure Arch for the cranial caudal direction III Rotate the knobs on each side of the SBRT Secure Arch until the clamps properly fasten onto the baseplate Do not apply too much f...

Страница 7: ...lock the Plate into position 5 3 PLACING THE ABDOMINAL VACUUM CUSHION I Slot the white pins of the cushion into the holes at the top of the SBRT Secure Arch and slide the white sliding mechanism into...

Страница 8: ...100 CouchStrips and Sidepieces II Placing the OmniBoard Align the Omniboard with the CouchStrips and slot it in place Make sure that no movement is possible between the baseplate and the couch 6 2 ATT...

Страница 9: ...ETACHING THE MULTIBOARD HXP RAIL SYSTEM The MultiBoard HXP Rail System can be attached to the MultiBoard HXP baseplate Align the bottom of the MultiBoard HXP baseplate with the rail system and slot it...

Страница 10: ...he device in a clean ventilated environment Dispose of the device in accordance with the local laws and regulations applicable to your country 8 Warranty The MacroMedics products are warranted to be f...

Страница 11: ...se Caution Indicates the need for the user to consult the instructions for use for important cautionary information such as warnings and precautions that cannot for a variety of reasons be presented o...

Страница 12: ...MacroMedics BV Oostbaan 670 2841ML Moordrecht The Netherlands 31 0 182 389 777 info macromedics com www macromedics com Visit www macromedics com for a digital copy of this manual...

Отзывы: