Manual QUANTOFIX
®
Relax
93
MN
Manual QUANTOFIX
®
Relax EN V 1.10 / 01.14
17. Error messages
If you cannot solve a problem with the instructions in the table, please contact
your distributor or MACHEREY-NAGEL directly (see 18.5 Contact, p. 95).
Error messages are displayed as cleartext and are self explanatory.
Error Message
Source
Solution
“Remove strip”
Test strip has not been
removed from slide af-
ter measurement in the
auto start mode
Remove test strip and
start new measurement
“Wrong position”
Strip has not been
pushed to the far end of
the slide
Start a new measure-
ment and push the strip
to the end of the slide
“No paper”
There is no paper left or
the printer cover is still
open
Insert new paper and
close printer cover
“Battery empty”
Empty batteries
Change batteries or
connect to an outlet.
“Instrument doesn’t
start”
No power or power con-
nection broken
Check if all connections
are correctly set and if
the outlet is working.