Timer (S0050)- Specifications and Terminal Connections

The timer is an Elapsed Time Indicator that is powered by an internal 3 volt lithium battery.

 

It comes pre-assembled in its housing and ready for mounting to the sink.

 

Specification

Description

Start/Stop Input

NPN, Contact Closure; Accumulates time when connected to common; Low State: 

< 1.0 VDC, High State: > 2.0 VDC (28VDC max)

Security Input

Allows access to panel reset and programming features

Remote Reset Input

 NPN or Contact Closure to common; level sensitive

Power Source

Single or dual 3V Lithium battery; typical 5 years life w/single battery, 10 years 

w/dual batteries

Ranges & Resolution

Seconds, minutes to 1/10, hours to 1/10,hours: minutes: seconds

Display

12 mm high, Supertwist LCD; 7 digits; “Low Bat” indicator

Backlighting

Green Illumination over whole viewable area. Requires 10 to 28 VDC power source

Dimensions & Mounting See dimensions picture below. Panel Mount with supplied mounting bracket and 

gasket

Connections

Screw terminals

Weight

Approximately 64 grams (2.25 ounces)

Rear Terminal Connections

Dimensions of Timer

5

6

7

8

4

3

2

1

72mm (2.83")

68mm (2.66")

33mm

(1.30")

36mm

(1.42")

5. Front Panel Program Enable - Allows access to

program mode when tied to common

6. Not Used

7. Not Used

8. DC Supply Input - For backlighting

1. Common

2. Not Used

3. Timing Enable - NPN or dry

contact inputs

4. Remote Reset - Resets time value

when tied to common

5

6

7

8

4

3

2

1

+

-

10 - 28 VDC

Содержание S0202-EST-KT

Страница 1: ...d and understand all of the instructions and safety information in this manual before operating this product MAN 028 Rev B 08 2021 2021 MAC Medical Inc Add on Scrub Sink Timer Kit for Installation on Economy Processing Sinks ...

Страница 2: ...llation and Timer setup programming instructions for the devices contained in this kit Table of Contents Parts List for S0202 EST KT 3 Plumbing Installation 4 Timer S0050 Specifications and Terminal Connections 7 Terminal Wiring Connections and Battery Replacement 8 Timer Operation Display and Programming 9 InLine Flow Switch Timer Controller S0130 01 Specifications Switch Rating and Wiring Diagra...

Страница 3: ...tion QTY 1 S0050 Timer Eagle 1 2 S0130 01 Inline Flow Switch Timer Control Brass 1 3 S0090 01 Timer Housing Weldment 1 4 S0019 Union 1 2 Copper 3 5 S0049 Elbow 1 2 Copper 90 4 6 S0165 Pipe Fitting Reducer 1 2 Sweat X 1 4 Npt Male 2 7 S0069 1 2 Copper Hard Type L Cierro Brand 10 Links Only 3 ft ...

Страница 4: ... inches below elbow see red cut line New Unions New Union Flow Switch Timer After the cuts the original pipe will be relocated on the sink The sketch at left shows the new plumbing configuration outlined in black and showing unions and timer location In the new plumbing configuration use the kit plumbing items flow switch timer two reducers three unions four elbows additional pipe length to rerout...

Страница 5: ...iginal plumbing straightness S0049 Elbows S0165 Reducers top and side S0130 01 Timer S0019 Unions S0049 Elbow S0019 Union S0049 Elbow The cut original pipe is now located to the left from its previous position 18 1 2 length pipe S0069 or trim to allow for elbow 2 1 4 Length S0069 pipe 1 1 2 length pipes S0069 5 length pipe S0069 Detail See Detail inset below left ...

Страница 6: ...in the illustration below Two holes are for mounting the timer housing one for threading the three flow switch timer wires to the top side for connection to the timer read out Place a grommet in the wire threading hole Grommet to be supplied by others Mount the timer housing to the top of the sink in the 0 31 diameter drilled holes Mounting studs are provided on the housing Secure to sink top with...

Страница 7: ...w single battery 10 years w dual batteries Ranges Resolution Seconds minutes to 1 10 hours to 1 10 hours minutes seconds Display 12 mm high Supertwist LCD 7 digits Low Bat indicator Backlighting Green Illumination over whole viewable area Requires 10 to 28 VDC power source Dimensions Mounting See dimensions picture below Panel Mount with supplied mounting bracket and gasket Connections Screw termi...

Страница 8: ...be run on two batteries to extend the battery life to 10 years Once the battery is in place the unit will go into a self test mode and all the segments on the LCD display will be illuminated The self test mode is exited by depressing the Next key which will then display the model number 6 Depress the Next key again to ready the unit for operation Terminal Wiring Connections and Battery Replacement...

Страница 9: ...ing parameters can be accessed when the Program Enable input is active by pressing the Down key To edit a parameter use the Down key to scroll until the desired parameter appears on the screen Pressing the Next key will cause the leftmost digit of that value to begin to flash Use the Next and Down keys in combination to choose individual digits and change their value Front Panel Reset Enable When ...

Страница 10: ...h Rating and Wiring Diagram This is an explosion proof brass flow switch actuation set point 0 50 GPM 1 89 LPM and calibrated for water at standard conditions It is used for accurate detection of excessive or insufficient flow rates Switch Ratings Max Resistive Load This unit was calibrated in a vertical position with lead wires up Install unit in piping system using standard pipe fitting procedur...

Страница 11: ...ally be removed from these switches Occasional wipe down cleaning when excessive contamination is present is all that is normally required To Clean Remove unit from system and disassemble as shown below Clean all parts reassemble and reinstall unit Note 50 micron filtration is recommended Piston Inlet Fitting Housing Spring O Ring ...

Страница 12: ...ll Free 877 828 9975 www macmedical com MAC Medical Inc Printed in USA Publication No MAN 028 Rev B August 2021 Information regarding this product is subject to change without prior notice Customer needs are our first priority To place an order contact our customer service department at 618 476 3550 or 877 828 9975 or by email at sales macmedical com ...

Отзывы: