Operating Your  Nebulizer

Follow the previous cleaning instructions prior
to using your Nebulizer for the first time or
after it has been stored for an extended 
period. 
Remember: Always unplug the Nebulizer and
turn off the switch at the bottom before clean-
ing and filling and after each use. 

To fill the water reservoir,

1.Remove the cover by pulling

straight up, 

Fig. 1.

Remove the medicine cup 
if applicable.

2.Select either tap or distilled

water* that has a tempera-
ture range of 68ºF - 86ºF
(20ºC - 30ºC), 

Fig. 2

*Distilled water is preferred.

3.Fill the water reservoir with

tap or distilled water to the
indicated level (between the
2 lines located inside the
water reservoir), 

Fig. 3.

Remember: Do not use solutions other  than
water in the reservoir. It may damage the
transducer — Fill to the correct level.
NOTE: 
• The use of cool water (below 68ºF or 20ºC) 

will result in a lower rate of nebulization.

• The use of warm water (above 86ºF or 

30ºC) will result in a higher rate of 
nebulization.

To fill the medicine cup,

1.  Place the medicine cup inside the water

reservoir. 

2.  Pour the medicine directly into the 

medicine cup.

3.  Replace the cover.

Remember: 
• The maximum capacity of the medicine

cup is 8 ml. and should not be overfilled.

• Dilute dense medications or medication 

volumes 2 ml. or less with distilled water or 
saline solution to obtain a volume of at least 
4 ml.

• Only use medications under the instructions 

of your physician.

Using the AC adapter or 
battery pack

Your Nebulizer operates by use of the AC
adapter or battery pack, both included with
your Nebulizer.

Using the AC adapter,

1) Insert the adapter plug into

the unit’s AC port, 

Fig 4.

2) Plug the adapter into a

standard AC outlet.

NOTE: It is not necessary to install the bat-
tery pack when using AC voltage. However,
you may wish to install the battery pack for
stability and storage purposes.

Using the battery pack,

Your Ultrasonic Nebulizer can operate with
the use of a rechargeable battery when AC
power is not available. The life of the battery
is approximately 2 years. However, battery
life will vary depending on
usage, storage and charging
conditions.

1)  Install the battery pack by

sliding it into the bottom
right side of the Nebulizer,
Fig. 5.

2)  Flip the switch at the bot-

tom to the “OFF/CHARGE”
position.

3)  Connect the adapter plug

to the unit as described
above, “Using the AC
adapter”.

4) To charge the battery, plug the adapter

into a standard AC outlet, based on the
following:
Charging Time
1st Time use:

24 hours

Thereafter:

4-12 hours

5) Unplug the adapter after the battery is

fully charged.

IMPORTANT:

For first time use or after storing the bat-
tery for a period of time, charge the bat-
tery at least 24 hours. Thereafter, it is
only necessary to charge 4 to 12 hours
depending on the length of usage.

A fully charged battery will last for 20-30
minutes of continuous use.

Fig. 1

Fig. 4

Fig. 2

Fig. 3

5

Fill
lines

AC port
(bottom of unit)

Fig. 5

Содержание 40-270-000 MABISMist II

Страница 1: ...ed for use in the treatment of asthma COPD and other respiratory ailments in which an aerosolized medication is required during therapy The MABISMist II Ultrasonic Nebulizer is not intended for use with Pentamidine It is intended for single patient use Federal U S A law restricts this device to sale by or on the order of a licensed physician English Spanish TOLL FREE CUSTOMER CARE HELPLINE 1 800 6...

Страница 2: ...ar use or purpose whether or not the use or purpose has been disclosed to MABIS Healthcare in specifications drawings or otherwise and whether or not MABIS Healthcare s products are specifically designed and or manufactured by MABIS Healthcare for the buyer s use or purposes except for the limited warranty stated above MABIS will not be responsible for any indirect incidental special consequential...

Страница 3: ... unit Do not attempt to remove the transfer cone located in the base of the water reservoir If any abnormality occurs discontinue use until the unit has been examined and repaired The unit should not be left unattended while plugged in Do not use solutions other than water in the reservoir It may damage the transducer Do not tilt or shake the unit when in operation Opening the cover during operati...

Страница 4: ...high humidity Avoid placing the battery on metal or wet surfaces or inside a metal box This may short circuit the battery s positive and negative contact causing the battery to heat up or explode Cleaning Procedures It is important to clean your Nebulizer before the first time it is used and after each use Remember Disconnect the unit from the electrical outlet and never immerse the unit in water ...

Страница 5: ...east 4 ml Only use medications under the instructions of your physician Using the AC adapter or battery pack Your Nebulizer operates by use of the AC adapter or battery pack both included with your Nebulizer Using the AC adapter 1 Insert the adapter plug into the unit s AC port Fig 4 2 Plug the adapter into a standard AC outlet NOTE It is not necessary to install the bat tery pack when using AC vo...

Страница 6: ...w settings High Medium and Low 2 Based on instructions from your physician and or respiratory therapist press the airflow button to adjust the airflow volume The Airflow Volume Indicator Lights will show the level high medium or low selected Fig 10 To begin a treatment 1 Follow the instructions provided by your physician and or respiratory therapist 2 Your Nebulizer is designed to start at a lower...

Страница 7: ...firmly and properly properly mounted to the motor shaft A flashing red This is normal see Using the light on the Battery Pack The flashing bottom lamp light indicates proper AC when the AC connection and will change to adapter is green when the battery is connected fully charged The top light The battery needs to be changed from charged Nebulization will green to red decrease and then cease during...

Страница 8: ... Forest IL 60045 USA Toll Free Customer Care Helpline 800 622 4714 www mabis net E mail nebinfo mabis net Phone 847 680 6811 Fax 847 680 9646 8 Unit Specifications Name and model number MABISMist II Ultrasonic Nebulizer 40 270 000 Power 100 120 VAC or 220 240 VAC 50 60 Hz Power Consumption Approximately 10W Ultrasonic Frequency 2 5 MHz Nebulization 3 speed adjustable Partical Size 5 m MMAD Medicin...

Отзывы: