CALIBRATION

This unit is calibrated at the factory to meet the ordered 
specifications, therefore you usually do not need any cali-
bration.
For matching the signal to a receiving instrument or in case 
of regular calibration, adjust the output as explained in the 
following.

 

ZERO/SPAN & UP/DOWN SWITCHES

Fine zero and span adjustments can be done with the ZERO/
SPAN Selector and UP/DOWN switches.  Calibrated values 
are stored in the non-volatile memory, which will not be lost 
even when you turn off power supply to the unit.
The output 1 and output 2 can be independently adjustable.  
Zero and span are respectively set to 0% and 100% at the 
factory, and field calibration can be reset to ex-factory set-
ting if necessary.

Adjustable range: 

±5%

Zero/Span Selector

 

ZERO

:  UP/DOWN switches usable for zero adjustment.

 

OFF

:  

UP/DOWN switches unavailable.

 

SPAN

:  UP/DOWN switches usable for span adjustment.

UP/DOWN Switches

 

UP

:   

Pressing UP increases adjusted values.

 

DOWN

:  Pressing DOWN decreases adjusted values.

 

FRONT VIEW

■ 

FRONT VIEW

■ 

SIDE VIEW

■ 

BOTTOM VIEW

Zero/Span Selector

UP Switch

Status Indicator LED

Configurator Jack

DOWN Switch

Output Range Selector

V

I

S

Z

I

OFF

Z

S

V

 

BOTTOM VIEW

■ 

FRONT VIEW

■ 

SIDE VIEW

■ 

BOTTOM VIEW

Zero/Span Selector

UP Switch

Status Indicator LED

Configurator Jack

DOWN Switch

Output Range Selector

V

I

S

Z

I

OFF

Z

S

V

Zero/Span Selector

MODE

SW POSITION

SPAN

S

OFF

ZERO

Z

 

HOW TO CALIBRATE THE ZERO & SPAN

Use a signal source and measuring instruments of sufficient 
accuracy level.  Turn the power supply on and warm up for 
more than 10 minutes.
The output 1 and output 2 adjustments are linked except 
for (2) and (6).

1) Apply 0% input.

2) Select the Output 1 Zero: Slide the Zero/Span Selector to 

the position Z. 

 

Select the Output 2 Zero: Slide the Zero/Span Selector to 

the position Z while pressing UP switch.

3) Press UP or DOWN switch and adjust the output to 0%.  

Incrementing speed will be doubled when you keep press-

ing a switch.

4) Put the Zero/Span Selector back to the position OFF.

5) Apply 100% input.

6) Select the Output 1 Span: Slide the Zero/Span Selector to 

the position S. 

 

Select the Output 2 Span: Slide the Zero/Span Selector to 

the position S while pressing UP switch.

7) Press UP or DOWN switch and adjust the output to 

100%.  Incrementing speed will be doubled when you 

keep pressing a switch.

8) Put the Zero/Span Selector back to the position OFF.

9) Check Zero and Span adjustments again with 0% and 

100% input respectively.

10) If the output value is changed, repeat the above proce-

dure (1) – (9).

 

HOW TO RESET

After you calibrated manually with these switches, you 
can reset them to its ex-factory state by pressing both UP/
DOWN switches at once.

Содержание M8XM2

Страница 1: ...anel system The user may have to review the CE requirements in regard to the whole system and employ additional protective measures to ensure the CE conform ity Install lightning surge protectors for those wires connect ed to remote locations POWER INPUT RATING OPERATIONAL RANGE Locate the power input rating marked on the product and confirm its operational range as indicated below 24 V DC rating ...

Страница 2: ...e correctly con nected according to the connection diagram 2 DIP switches Check the output range selector setting 3 Power input Check the power input voltage 4 Input Check voltage across the terminal B C with a voltmeter to show 0 V at the minimum potenti ometer input 0 Ω and the same voltage as that across A C at the maximum input If the potentiometer wires are broken the output goes over 100 bel...

Страница 3: ...IEW SIDE VIEW BOTTOM VIEW Zero Span Selector UP Switch Status Indicator LED Configurator Jack DOWN Switch Output Range Selector V I S Z I OFF Z S V Zero Span Selector MODE SW POSITION SPAN S OFF ZERO Z HOW TO CALIBRATE THE ZERO SPAN Use a signal source and measuring instruments of sufficient accuracy level Turn the power supply on and warm up for more than 10 minutes The output 1 and output 2 adju...

Страница 4: ...ration Mode you can reset the calibrated status to its ex factory state by pressing both UP DOWN switches at once Confirm that the LED turns off when you press these switches I O RANGE MODIFICATION Configurator Software is used to change input types and ranges For changing output ranges set the Output Range Selector and then select by Configurator Software FRONT VIEW SIDE VIEW BOTTOM VIEW Zero Spa...

Отзывы: